Chapter 4 — Adjustment
4-8 Level Detector Tables and Level Flatness Adjustment
MG3702xA MM
PN: 10370-10372 Rev. B
4-7
4-8
Level Detector Tables and Level Flatness Adjustment
These adjustments update the high, mid, and low gain detector tables along with the level flatness tables.
These adjustments require use of the Levels Calibration Station as described below.
Required Equipment:
•
PC with GPIB installed
•
Level Cal Software, p/n 57877, revision 2.00 or higher
•
Anritsu ML2438A Power Meter
•
Anritsu MA2474D Power Sensor
•
Anritsu 41KC-10 10 dB Attenuator
•
ML2530A Calibration Receiver
•
T2579 Mixer Fixture
•
Mini-Circuits Filter, Model BLP-1.9, Qty 2
•
MG3692C synthesizer
Equipment Setup
Connect the equipment as shown in
Procedure
1.
Power on the MG3702XA and allow the unit to warm up for 1 hour.
2.
Connect the GPIB from the Level Cal Station to the MG3702xA.
3.
Open the AMS Test Executive software.
Figure 4-2.
Levels Calibration Station Setup
Connect to DUT RF Output
as directed by procedure.
DUT RF Output
Mini-Circuits Filters
Model BLP-1.9
High Band
In/Out
Control
Low Band
In/Out
Control
Connect to DUT RF Output
as directed by procedure.
10 dB
Attenuator
Connect to DUT RF Output
as directed by procedure.
GPIB
PC
ML2530A
ML2438A
MA2474D
Mixer
MG3692C
MG3702xA
DUT
T2579
DUT
LO
To
ML2530
ML2530A
Calibration Receiver
MG37022A
Signal
Generator
Standby
Operate
10 MHz
REF IN
10 MHz
REF OUT
10 MHz
REF IN
Summary of Contents for MG3702xA Series
Page 4: ......
Page 5: ......
Page 8: ...TG 8 PN 10370 10372 Rev B MG3702xA MM ...
Page 24: ...1 12 Test Equipment List Chapter 1 General Information 1 10 PN 10370 10372 Rev B MG3702xA MM ...
Page 42: ...2 5 RF Deck Assemblies Chapter 2 Functional Description 2 18 PN 10370 10372 Rev B MG3702xA MM ...
Page 120: ...5 5 Troubleshooting Tables Chapter 5 Troubleshooting 5 24 PN 10370 10372 Rev B MG3702xA MM ...
Page 182: ...Index 4 PN 10370 10372 Rev B MG3702xA MM ...
Page 183: ......