4-7 Frequency Reference Adjustment
Chapter 4 — Adjustment
4-6
PN: 10370-10372 Rev. B
MG3702xA MM
4-7
Frequency Reference Adjustment
This adjustment improves the frequency accuracy of the MG3702XA.
Required Equipment:
•
Absolute Time Corp. Model 300, 10 MHz Frequency Reference
Equipment Setup
Connect the equipment as shown in
Adjust the Frequency Reference as follows:
1.
Power on the MG3702XA and allow the unit to warm up for 1 hour.
2.
Preset the unit by pressing the
Preset
hard key.
3.
Press
System
| Maint |
12345
|
Enter
| Calibration | Frequency.
4.
Press
Reference Cal.
5.
Connect the
10 MHz Output from the Frequency Reference
to the
10 MHz REF IN connector
at the
rear panel of the MG3702XA.
6.
Press
Run Cal | Next
to continue.
The values are displayed on the screen.
7.
Press
Next.
8.
Press
Save to Nonvolatile.
9.
Choose
Yes
to overwrite existing values.
10.
The adjustment is complete. Record as completed in the test record.
Figure 4-1.
Frequency Reference Adjustment
Summary of Contents for MG3702xA Series
Page 4: ......
Page 5: ......
Page 8: ...TG 8 PN 10370 10372 Rev B MG3702xA MM ...
Page 24: ...1 12 Test Equipment List Chapter 1 General Information 1 10 PN 10370 10372 Rev B MG3702xA MM ...
Page 42: ...2 5 RF Deck Assemblies Chapter 2 Functional Description 2 18 PN 10370 10372 Rev B MG3702xA MM ...
Page 120: ...5 5 Troubleshooting Tables Chapter 5 Troubleshooting 5 24 PN 10370 10372 Rev B MG3702xA MM ...
Page 182: ...Index 4 PN 10370 10372 Rev B MG3702xA MM ...
Page 183: ......