Selecting a
Sweep
Frequency/Step
Power Mode
In analog sweep frequency/step power mode or step
sweep frequency/step power mode, a power level
step occurs after each frequency sweep. The power
level remains constant for the length of time re-
quired to complete each frequency sweep. Available
menus let you control the type of power level sweep
(linear or logarithmic), the power level sweep range,
and the step size.
To select an analog sweep frequency/step power
mode, start with an analog sweep menu display; to
select a step sweep frequency/step power mode, start
with a step sweep menu display. Then press
Level
.
At the resulting menu display, press
Level Sweep
.
The Level Sweep menu (below) is displayed.
This menu lets you perform the following:
q
Select a power level sweep range (edit the
sweep start and stop power level parameters)
q
Set the step size
q
Set the number of steps
q
Access the Level Sweep Trigger menu
NOTE
To select logarithmic or linear power level
sweep or to select a power level sweep range, re-
fer to the procedures on pages 3-59 and 3-60.
MG369XB OM
3-61
Local (Front Panel) Operation
Power Level Sweep Operation