Programming with GPIB Commands
2-22 Miscellaneous Commands
MG369xC GPIB PM
PN: 10370-10374 Rev. F
2-61
2-22 Miscellaneous Commands
is a list of miscellaneous command mnemonic codes that do not fit into any of the other
classifications. These commands provide the following operations:
•
GPIB Address Change
•
CW Ramp
•
Secure Mode
•
Returning the MG369xC to local control
•
Instrument Reset
•
Serial Number Entry
•
Phase offset control
Table 2-23.
Miscellaneous Commands
Mnemonic
Code
Function
ADD
Opens the GPIB address parameter.
Programming
Example:
Programming “
ADD 13 ADR
“ changes the instrument GPIB address to 13.
CS0
Turns off the CW ramp.
CS1
Turns on the CW ramp.
DS0
Turns on the secure mode.
DS1
Turns off the secure mode.
MR
Performs a master reset of the instrument by setting all of the instrument parameters to factory
default.
PS0
Turns off the Phase Offset.
PS1
Turns on the Phase Offset.
PSZ
Zeros the Phase Offset display.
RL
Returns the instrument to local (front panel) control.
RST
Resets the MG369xC to its default settings. See the note below.
SNR
Enter the instrument serial number.
Note
Sending the
RST
command clears the current instrument setup. If this setup is needed for future
testing, save it as a stored setup (
Section 2-12 “Stored Setup Commands” on page 2-34
) before
sending
RST
.
Summary of Contents for MG369 C Series