2-19 User Level Calibration Commands
Programming with GPIB Commands
2-50
PN: 10370-10374 Rev. F
MG369xC GPIB PM
2-19 User Level Calibration Commands
lists the user level calibration commands. These commands provide for (1) activating individual
user level calibration tables, (2) sending the tables to the controller, and (3) recalling the tables from the
controller.
The user level (flatness correction) calibration function provides for calibrating out path variations in a test
setup. This is accomplished by means of an entered power-offset table from a GPIB power meter or calculated
data. When the user level calibration table is activated, the set power level is delivered to the point in the test
setup where the calibration was performed. Up to five user level calibration tables from 2 to 801 frequency
points/table can be created and stored in MG369xC memory for recall. (Refer to “Leveling Operations” in
Chapter 3 of the MG369xC Operation Manual for user level calibration procedures. )
The commands,
LU1
thru
LU5
, each activate an individual user level calibration table (#1 thru #5). The
LU0
command turns off the active user level calibration table. The
LUS
command sends all five tables of user level
calibration data to the controller where they are stored in a binary data file. While stored in the file, the data
can be edited. The
LUR
command readies the MG369xC to receive the five tables of user level calibration data
from the controller.
shows an example program for saving and recalling user level
calibration tables.
Table 2-20.
User Level Calibration Commands
Mnemonic
Code
Function
LU0
Turns off the active user level calibration table.
LU1
Activates user level calibration table #1. Turns off any other active user level calibration table.
LU2
Activates user level calibration table #2. Turns off any other active user level calibration table.
LU3
Activates user level calibration table #3. Turns off any other active user level calibration table.
LU4
Activates user level calibration table #4. Turns off any other active user level calibration table.
LU5
Activates user level calibration table #5. Turns off any other active user level calibration table.
LUR
Readies the MG369xC to receive five tables of user level calibration data from the controller.
LUS
Sends all five tables of user level calibration data to the controller.