2-9 Modulation Commands
Programming with GPIB Commands
2-26
PN: 10370-10374 Rev. F
MG369xC GPIB PM
Stepped Delay Mode
The stepped delay mode automatically increments or decrements the pulse delay 1 value according to step
delay parameters. Stepped delay mode is only available when the triggering commands
PTG3
(delayed) or
PTG5
(triggered with delay) are specified. The
SD1
command turns on the stepped delay mode. The pulse delay 1
start time can be set using the
SDS
parameter entry command; the delay 1 stop time can be set using the
SDE
parameter entry command. Use the
SDD
parameter entry command to set the step size and the
SDL
parameter
entry command to set the dwell-time-per-step. The
SD0
command turns off the stepped delay mode.
Programming Notes:
Use the
PTR
and
PTF
commands in conjunction with the pulse trigger commands
PTG2
(gated),
PTG4
(triggered), and
PTG5
(triggered with delay) to specify whether the pulse generator is triggered by the rising or
falling edge of the external trigger pulse.
External Pulse Modulation Function
The
XP
command turns on the external pulse modulation function. The
PO
(or
SW0
) command turns off the
pulse modulation function.
Modulation Command List
Table 2-9.
Modulation Command List (1 of 2)
Mnemonic
Code
Function
AM0
Turns off the internal or external AM function.
AM1
Turns on the external AM function in Linear mode.
AM2
Turns on the external AM function in Log mode.
AM7
Turns on the internal AM function in Linear mode.
AM8
Turns on the internal AM function in Log mode.
AMO
Turns off the internal AM function generator if it is on.
AMW(x)
Selects the internal AM waveform, where x = 1-sine wave, 2-square wave, 3-positive ramp,
4-negative ramp, 5-Gaussian noise, 6-uniform noise, 7-triangle wave, 8-user defined.
DPT
Internal triggered pulse with delay.
EP0
Selects TTL-low to turn RF on during pulse modulation.
EP1
Selects TTL-high to turn RF on during pulse modulation.
FM0
Turns off the internal or external FM function.
FM1
Turns on the external FM function in Unlocked Narrow mode.
FM2
Turns on the external FM function in Locked Low-Noise mode.
FM7
Turns on the internal FM function in Unlocked Narrow mode.
FM8
Turns on the internal FM function in Unlocked Wide mode.
FM9
Turns on the internal FM function in Locked mode.
FML
Turns on the external FM function in Locked mode.
FMN
Turns on the internal FM function in Locked Low-Noise mode.
FMU
Turns on the external FM function in Unlocked Narrow mode (Same as FM1).
FMW
Turns on the external FM function in Unlocked Wide mode.
FWV(x)
Selects the internal FM waveform, where x = 1-sine wave, 2-square wave, 3-positive ramp,
4-negative ramp, 5-Gaussian noise, 6-uniform noise, 7-triangle wave.
GP
Internal gated pulse on.