2-6 Analog and Step Sweep Commands
Programming with GPIB Commands
2-16
PN: 10370-10374 Rev. F
MG369xC GPIB PM
Analog/Step Sweep Select
Five commands are available—
SWP
selects an analog sweep,
SSP
and
LIS
select a linear step sweep,
LGS
selects a logarithmic step sweep, and
MAN
selects a manual (step) sweep. The selected sweep mode applies to all
sweep ranges.
shows an example of a sweep command string.
Programming Notes:
Commanding either
SWP
or
SSP
does not, by itself, provide a swept-frequency output. It only determines
whether the swept-frequency output will be an analog or step sweep. If, on the other hand, a frequency sweep
is being output by the signal generator when one of these commands is received, that sweep will assume the
commanded sweep mode.
The
MAN
command only provides for setting up the MG369xC for a manual sweep. It must be accompanied by
the
RL
command to return the instrument to local (front panel) control in order for the operator to perform the
manual sweep.
Figure 2-3.
Example of a Sweep Command String
SF1
Selects the F1-F2 sweep range.
Selects Auto triggering.
Selects analog sweep.
Selects the Single triggering mode.
Triggers a single sweep.
Selects linear step sweep.
Triggers a single sweep.
AUT
SWP
EXT
TRS
SSP
TRS