Making Measurements
Using the Power Sensor
3-14
PN: 10585-00016, Rev. A
MA24104A UG
Table 3-1, describes the number of averages needed to attain a certain noise level for a particular power level
measurement with the Low Aperture Time mode setting.
Table 3-2, describes the number of averages needed to attain a certain noise level for a particular power level
measurement with the High Aperture Time mode setting.
Note: The values in the following tables are typical and should be used as a reference only.
Table 3-1.
MA24104A Averaging Table (Low Aperture Time, Default Mode)
Input Power
(dBm)
Input Power
(W)
Number of
Averages
Needed for
<
±
0.20 dB
Noise
Number of
Averages
Needed for
<
±
0.15 dB
Noise
Number of
Averages
Needed for
<
±
0.10 dB
Noise
Number of
Averages
Needed for
<
±
0.05 dB
Noise
Number of
Averages
Needed for
<
±
0.01 dB
Noise
50
100
1
1
1
1
2
45
31.6
1
1
1
4
16
40
10.0
1
1
1
20
78
35
3.16
1
1
1
1
1
30
1.00
1
1
1
1
1
25
0.316
1
1
1
1
7
20
0.100
1
1
1
3
61
15
0.0316
2
3
7
25
–
10
0.0100
16
28
62
245
–
5
0.00316
158
–
–
–
–
Table 3-2.
MA24104A Averaging Table (High Aperture Time)
Input Power
(dBm)
Input Power
(W)
Number of
Averages
Needed for
<
±
0.20 dB
Noise
Number of
Averages
Needed for
<
±
0.15 dB
Noise
Number of
Averages
Needed for
<
±
0.10 dB
Noise
Number of
Averages
Needed for
<
±
0.05 dB
Noise
Number of
Averages
Needed for
<
±
0.01 dB
Noise
50
100
1
1
1
1
1
45
31.6
1
1
1
1
1
40
10.0
1
1
1
2
5
35
3.16
1
1
1
1
1
30
1.00
1
1
1
1
1
25
0.316
1
1
1
1
1
20
0.100
1
1
1
1
4
15
0.0316
1
1
1
2
38
10
0.0100
1
2
4
16
–
5
0.00316
10
18
39
153
–