5.8 Setting Sync Output
5-27
5
Generating Pulse Signal
The Sync Output amplitude cannot be set.
The time interval outputting the pulse varies depending on the pattern
length and bit rate when Sync Output is PPG Pattern Sync .
Select the pattern for the measurement referring to Table 5.8-1.
Table 5.8-1 Cycle of Sync Output (PPG Pattern Sync)
Pattern Name
Bit Rate (kbit/s)
155520 622080 1250000
2488320
4250000
10000000
PRBS 2^7–1
52.3
s
13.1
s
6.5
s
3.27
s
1.91
s 812
ns
PRBS 2^9–1
210
s
52.6
s
26.2
s
13.1
s
7.70
s 3.27
s
PRBS 2^15–1
13.48 m s
3.37 ms
1.68 ms
843
s
493
s 210
s
PRBS 2^23
–
1
3.45 s
863 ms
429 ms
216 ms
126 ms
53.7 ms
PRBS 2^31–1
884 s
221 s
110 s
55.2 s
32.3 s
13.7 s
CAUTION
The impedance of the Sync Out connector is 50
Ω
.
Measurement may not be performed correctly if a
coaxial cable with another impedance is used or if the
impedance of the DUT is not 50
Ω
.
The output voltage of the Sync Out connector is –0.5 to
0 V. Check that the output voltage does not exceed the
DUT maximum input specifications. If it does, connect
an attenuator to the Sync Out connector.
Summary of Contents for BERTWave MP2100B
Page 24: ...xxiv...
Page 96: ...Chapter 2 Before Use 2 42...
Page 112: ...Chapter 3 Connecting with DUT 3 16...
Page 250: ...Chapter 7 Measuring Waveform 7 64 Figure 7 12 1 Marker Display...
Page 262: ...Chapter 7 Measuring Waveform 7 76...
Page 276: ...Chapter 8 Operating Optical Interface 8 14...
Page 304: ...Chapter 9 Performance Test 9 28...
Page 320: ...Chapter 10 Maintenance 10 16...
Page 374: ...Appendix E Performance Test Record Form E 10...
Page 382: ...Index Index 6...