Cleaning
the
water
filter
To clean the water filter (located on the water supply line), follow the procedure detailed below.
1.
Unscrew the filter from the instrument.
2.
Unscrew the filter clamp from the water supply line.
3.
Use compressed air to blow backwards through the
filter (blow opposite the arrow on the filter).
4.
Reconnect the filter and fittings.
5.
Turn on the water and check for leaks.
Replacing
the
Fuses
To replace the fuses in the ANKOM
XT10
Extraction System, follow the procedure detailed below.
1.
Turn off the instrument power and unplug the Power Cord from the outlet.
2.
Remove the two screws on the Top Cover.
3.
Swing the Top Cover down exposing the electrical compartment.
4.
Locate the Fuse Holders on the terminal strip. There
are two fuses within the instrument; one for each pole
(100 – 120V 10A or 220 – 240V 5A).
5.
Press the tabs to lift the Fuse Holders up.
6.
Replace the fuses and close the Fuse Holders completely.
7.
Close the Top Cover and secure it with the two screws that you previously removed.
Fuse
Holders
Tab
to
open
Fuse
Holder
Fuse
Filter