Replacing X59 Analog level sensor 

 and Fiber optic cable with glass tip 

XT15/XT15I

 

Service Procedure 

X5S012

 

JAD 4/2/2019               

 

 

ANKOM Technology 

 

2052 O’Neil Road 

 Email: [email protected]

 Phone: 315-986-8090 

 Fax: 315-986-8091 

pg. 1 of 4 

 

Note: Follow steps 1-8 for level sensor module replacement and 8-12 for Fiber optic cable replacement. (1-12 for 

both) 
 

1.

 

Unplug the Instrument. 

2.

 

Unscrew the four screws located on each side of the cabinet back. Lift the cabinet back while removing it from 
the instrument. 

3.

 

Open the electrical box by removing the two screws located on top of the instrument. 

4.

 

Loosen the two retainer screws using a flat head screwdriver (figure 1) 
 

                                                                                         

 

                                                                                                                                            Figure 1 

                                                                                         

 

                                                                                                                                              Figure 2 

5.

 

Gently pull fiber optic cable from level sensor. 
 
 
 
 
 

Summary of Contents for X59

Page 1: ...ensor module replacement and 8 12 for Fiber optic cable replacement 1 12 for both 1 Unplug the Instrument 2 Unscrew the four screws located on each side of the cabinet back Lift the cabinet back while removing it from the instrument 3 Open the electrical box by removing the two screws located on top of the instrument 4 Loosen the two retainer screws using a flat head screwdriver figure 1 Figure 1 ...

Page 2: ...ck Locate the cables that extends from the Level Sensor and cut any cable ties that bundle it with other cables Note This cable should contain 3 wires 3 Red 4 Black I1 white which attach to a terminal block Detach these from the terminal block by inserting and pushing the tip of a flat head screwdriver into the square Slot on the terminal block that is next to the wire you desire to remove This wi...

Page 3: ...52 O Neil Road Email service ankom com Phone 315 986 8090 Fax 315 986 8091 pg 3 of 4 7 Remove the two 3 32 hex head screws from the back side of the electrical cabinet This will free the level sensor 8 To install new Level sensor reverse steps 1 7 Note be sure to reinstall rubber gasket in between level sensor module and electrical cabinet ...

Page 4: ...sensor fitting from the back of the vessel using a 11 16 or adjustable wrench 11 To install new fiberoptic and glass tip reverse steps 10 8 along with 5 12 Once you have installed the new X59 Analog level sensor module and Fiberoptic cable with glass tip You are now ready for testing VERY IMPORTANT As soon as there is pressure in the vessel apply a soap solution to all threads on the vessel fittin...

Reviews: