Instruction Manual

Instrument Model Number 

Instrument Serial Number 

Form Number AIC2071 © 8/10

Revised: 8/10

Supersedes: NEW

"IZMSG" 

Electromagnetic Flowmeter

Summary of Contents for IZMSG

Page 1: ...156 Auriesville Road Fultonville NY 12072 1 800 833 0081 Fax 518 922 8997 Instruction Manual Instrument Model Number Instrument Serial Number Form Number AIC2071 8 10 Revised 8 10 Supersedes NEW IZMSG...

Page 2: ...handling The warranty is void if a customer modifies or services the flowmeter physically or electronically in any way Anderson Instrument makes no additional warranties expressed or implied whether...

Page 3: ...measuring principle used in the IZMSG flowmeter The instructions in this document address the preparation installation and start up of the IZMSG flowmeter meter body and converter as separate units In...

Page 4: ...attached to each end of the flow tube with a connection gasket and held in place by bolts The meter body incorporates two electromagnetic coils and a PFA lined flow tube in a stainless steel housing T...

Page 5: ...ing Electrical Connections for drawings and additional information about the cables The Electrode Signal Cable is a 3 conductor cable with a shield wire It is used to connect the electrodes in the met...

Page 6: ...cuit Board There are three circuit boards inside the converter 300 SMB1 Main Board Figure 1 3 300 SACPSB AC Power Supply Board Figure 1 4 and 300 SJB2 Junction Board Figure 1 5 In a converter with the...

Page 7: ...Anderson Instrument Company 1 5 AIC 2071 Figure 1 4 300 SACPSB AC Power Supply Circuit Board Figure 1 5 300 SJB2 Junction Circuit Board...

Page 8: ...1 6 IZMSG Compact Electromagnetic Flowmeter AIC 2071 Figure 1 6 300 SDKB Display and Keypad Circuit Board converter with LCD display...

Page 9: ...hes the meter body serial number The stainless steel connection adapters are factory assembled to the inlet and outlet sides of the meter body flow tube by bolts see Figure 2 1 The connection adapters...

Page 10: ...it passes through the flow tube In a vertical pipe section install the meter body in the upward direction of flow In a horizontal pipe section install the meter body at the lowest possible point of th...

Page 11: ...Anderson Instrument Company 2 3 AIC 2071 Guidelines for properly locating the meter body are summarized in the following table...

Page 12: ...sed in the previous subsection Install the meter body with a minimum of 5 pipe diameters of straight unobstructed pipe upstream from the meter body and 3 pipe diameters of straight unobstructed pipe d...

Page 13: ...roper clearance and support for the meter body with connection adapters See Figure 2 2 and the table below for meter body dimensions and weight 2 Place the meter body in line with the arrow on the met...

Page 14: ...h different serial numbers The flowmeter will malfunction or operate improperly Location Install the converter as close as possible to the meter body to avoid extraneous inductive pick up Do not subst...

Page 15: ...rter installation procedures may need to be tailored to the specific site To install the converter 1 Make sure the converter has the proper power supply voltage configuration for the specific installa...

Page 16: ...2 8 IZMSG Compact Electromagnetic Flowmeter AIC 2071 Figure 2 3 Converter Dimensions with without Mounting Brackets...

Page 17: ...ers See Figure 3 1 and following table for details Plug Each IZMSG meter body and converter is shipped with plugs and seal rings at all entry points When a plug is removed transfer its seal ring to th...

Page 18: ...Figure 3 2 Never enter the enclosure from the top or sides Create a generous drip leg before each entry point to collect condensate from within the conduit or cable After wiring fill any void with an...

Page 19: ...ficient wire gauge to provide zero resistance to the actual earth ground connection point Connection Cables Anderson Instrument Company provides two factory ready connection cables with the flowmeter...

Page 20: ...using water tight connectors at the meter body and converter enclosure If desired both cables may be routed in the same conduit Separate the cables into two individual conduits adjacent to the meter b...

Page 21: ...able openings in the converter for the connection cables See Wire Openings and Connectors in the preceding subsection If conduit is used the coil drive cable and electrode signal cable may be run in a...

Page 22: ...to the terminal locations inside the meter body See the following table Cable Wire Meter Body Terminal Gray Electrode Signal Cable red 14 white 16 black 18 shield 13 Black Coil Drive Cable white 11 bl...

Page 23: ...a 1 4 inch crimp connector spade terminal to the coil drive cable shield wire 4 Loosen Main Board attachment lug screw located in the lower left corner of the converter Slide the spade terminal betwe...

Page 24: ...3 8 IZMSG Compact Electromagnetic Flowmeter AIC2071 Figure 3 8 Meter Body Converter Connections...

Page 25: ...d cable to the length of cable required Discard the factory prepared end 2 Remove 1 5 of external insulation and foil Retain the shield wire Note A factory supplied 1 4 spade terminal will be crimped...

Page 26: ...umentation Do not select an AC electrical supply circuit that is used for powering large machines or motors 2 Route the AC power supply wire through conduit in accordance with local regulations Use a...

Page 27: ...al block XP at the location marked PE In most cases the green safety ground wire in 115 VAC 3 wire service will provide a path to earth ground Caution Failure to connect an earth ground to the termina...

Page 28: ...n Board see Figure 3 9 and table 300 SJB2 Board Function D5 D6 D7 D8 LED indicators for digital outputs D11 D14 D17 D20 LED indicators for digital inputs S5 ABORT pushbutton reset S4 ZERO ADJUST pushb...

Page 29: ...and random FLASH under no flow conditions The maximum frequency signal output is 500 HZ Signal configurations of 1000 HZ are available through Anderson Instrument as a special order The open collector...

Page 30: ...2071 P Analog Output Figure 3 18 P 24 VDC Pulse Figure 3 19 Signal wires to external receivers are not to be bundled or to share an electrical conduit with AC power wires Use a separate conduit for al...

Page 31: ...rational parameters necessary to meet the customer s specific application requirements The pre programmed data is retained in EEPROM memory in the converter Therefore the only on site adjustment neces...

Page 32: ...erter with an LCD display or a portable MSD Service Terminal flowmeter parameters and other information such as the scaled digital pulse output MAX flowrate and calibration constants may be displayed...

Page 33: ...onditions the LED may flash so infrequently or briefly that the flash will not be visible During high flowrate conditions the LED may flash so rapidly that the LED appears to be continuously on D6 24...

Page 34: ...o the flowmeter as long as the conditions of the installation remain unchanged Once initiated by the Zero Adjustment pushbutton the hydraulic zero adjustment is done automatically within 40 seconds Du...

Page 35: ...temperature In order to maintain thermal stability close but do not tighten the converter cover during the hydraulic zero adjustment procedure except when access to the converter is necessary to pres...

Page 36: ...age 3083 recheck all required conditions for the hydraulic zero adjustment procedure Return to Step 5 and repeat the procedure 7 Confirm the successful hydraulic zero adjustment by visually inspecting...

Page 37: ...u only want to perform a Zero Test The Zero Adjust procedure will cause a new zero value to be stored in the non volatile memory of the converter thus changing the condition of the installation To a c...

Page 38: ...onds showing the progress of the automatic zero measurement At the end of the 40 second automatic zero measurement period the visual display shows the following Where XXX XX is the zero measurement ju...

Page 39: ...Position of CAL MEAS Switch Function Performed MEAS Zero Test CAL Zero Adjust Caution Before executing a Function 00 procedure check the position of the CAL MEAS switch to avoid performing an undesire...

Page 40: ...on Repeating Zero Test Measurement and Comparing Zero Adjust and Zero Test Values Reversing Direction of Positive Flow Measurement In certain installations it may be desirable to change the direction...

Page 41: ...n and modification of program parameters This allows the flowmeter to be programmed in the field to meet the exact conditions of the measuring application or for fine tuning the flowmeter to the contr...

Page 42: ...ave an integral LCD display and keypad The MSD Service Display consisting of a display and keypad that is functionally identical to the converter LCD display and keypad is packaged in a hand held encl...

Page 43: ...e used to move up and down a given menu or left and right to the neighboring menus on either side The default menu designated by Menu 00 is the Measuring Register The Measuring Register contains the d...

Page 44: ...5 4 IZMSG Electromagnetic Flowmeter Installation Instructions AIC 2071 Figure 5 2 Parameter Map IZMSG Compact Electromagnetic Flowmeter User Software Version V1 XX...

Page 45: ...e CAL MEAS switch is in the CAL position P The value of the current parameter can be changed by pressing the key and then entering the desired value P Confirm the new value with the key Note that the...

Page 46: ...3 EEPROM plugs into connector X6 on the 300 SMB1 Main Board Should it become necessary to substitute a meter body or converter in the field and break up a matched pair consult the factory for the pro...

Page 47: ...to view Menu 3 Since Qmax is the first parameter displayed in Menu 3 it is not necessary to use the or keys to scroll to Qmax The screen will display the current value for Qmax 4 Press the key The va...

Page 48: ...d totalizer Q flow rate per hour or Q per minute see Table 5 4 Q typ for hour minute selection 02 menu 02 See table 5 2 Meter Parameters Pipe diameter DN x x Unit of measure unit x Low Flow Suppressio...

Page 49: ...s IMP1 IMP2 x x Time constant for damping analog output response tp3 x x Minimum on time required for activation it1 it4 of inputs IN1 IN4 x x 04 menu 04 See table 5 4 Display Parameters Selection of...

Page 50: ...ffset x x Field calibration constant m spe x Field calibration constant for forward flow p spe x Field calibration constant for backward flow n spe x Compensation constant for low flow rates b spe x C...

Page 51: ...unlocked by Function 98 Parameter checksum for non critical parameters nparasum Parameters that can be unlocked by Function 98 Meter body type sentyp Meter body number senno Shows position of CAL MEA...

Page 52: ...of the full scale flow rate Qmax Range 0 00 4 00 Default 2 00 output mode Determines the operation of the pulse outputs Range 0 10 Default 1 Mode of operation see Table 5 7 average Determines the numb...

Page 53: ...by 99 or by activating the digital input IN1 standby mode 1 standby not possible standby mode 2 standby is possible Range 0 2 Default 1 parameter mode Security function for the parameters which allows...

Page 54: ...nes the number of pulses per volume unit seen at output IMP2 Only inflow is possible at Output mode 1 5 Range Application specific tp1 tp3 tp1 Determines the maximum on time duration for the pulse out...

Page 55: ...f the decimal point for the display of totalizer V2 Range 0 6 Default 1 decimal point Selects period or comma as the display symbol for the decimal point Range or Default display mode Determines wheth...

Page 56: ...aluation Factor for the positive flow Range p spe 0 is not accepted Default 1 00000 n spe Separate Evaluation Factor for the negative flow Range n spe 0 suppresses measurement during reverse flow Defa...

Page 57: ...98 Read only value is based on the adjusted parameter data nparasum Display of checksum for non critical parameters Parameters that can be unlocked by Function 98 Read only value is based on the adjus...

Page 58: ...me duration programmable in msec via tp1and tp2 Off time depends on scaling and flow rate Set tpx 0 for pulse pause ratio 1 1 for IMPx Both pv1 Same function as Output Mode 1 except output IMP 1 pulse...

Page 59: ...lation diagrams Test the connection of each wire by gently tugging the wire with needle nose pliers to ensure that the wire is fully engaged in its terminal Make sure that AC power is available to the...

Page 60: ...A 2 IZMSG Compact Electromagnetic Flowmeter AIC 2071 Converter...

Page 61: ...Anderson Instrument Company A 3 AIC 2071...

Page 62: ...A 4 IZMSG Compact Electromagnetic Flowmeter AIC 2071 Meter Body...

Page 63: ...Anderson Instrument Company A 5 AIC 2071...

Page 64: ...power to the converter 2 Empty the flow tube completely Dry the inside of the flow tube 3 Disconnect the coil drive cable wires from terminals 11 and 12 in the meter body terminal box 4 Disconnect the...

Page 65: ...18 20M ohms D 12 and 18 20M ohms E 11 and 12 70 to 120 ohms Where means greater than M means million 6 If the meter body fails to meet any of the measured resistance values in the table above the mete...

Page 66: ...be no substantial difference between the current Zero Test value and the original stored Zero Adjust value If there is a gross difference of 75 to 100 units or more between the current Zero Test meas...

Page 67: ...ain totalizer V Certain error messages will be deleted automatically if the condition causing the error is corrected Error Message Number Error Diagnosis Corrective Action 901 a AC power failure or AC...

Page 68: ...current to meter body is outside operational range Consult factory 932 No coil drive current to the meter body Visual display message no current a Check terminals 11 and 12 at both meter body and con...

Page 69: ...nditions required for hydraulic zero adjustment 1 All connection cable terminations are tight dry and in correct locations 2 Flow tube is full of product 3 No flow or leakage Conduct hydraulic zero ad...

Page 70: ...D17 D20 D5 D6 D7 D8 all outputs off the digital outputs are switched off o o o o OUT1 On only output IMP1 is active on x o o o OUT2 On only output IMP2 is active on o x o o OUT3 On only output IMP3 is...

Page 71: ...AIC 2071 Curr1 on coil current is constantly at app 100 110 mA metered at terminal 11 12 Curr2 on value of the metered coil current must be almost the same as Curr1 on 0 3 mA but with the reverse pola...

Page 72: ...n the 1820 s by Michael Faraday an English physicist He experimented with magnets and found that if a conductor moves through a magnetic field a voltage is induced in the conductor Applied to the flow...

Page 73: ...ansmitted to the converter is directly proportional to the flow velocity The signal is independent of the fluid properties and operating characteristics such as product density viscosity conductivity...

Page 74: ...Maximum Product Temperature 325 F 160 C o o Maximum Inlet Pressure 154 psi absolute 11 bar Minimum Inlet Pressure 1 4 psi absolute 0 1 bar Minimum Fluid Conductivity Standard 100 micromhoes cm Special...

Page 75: ...puts 1 suspend measuring with coil power supply off 1 remote reset internal totalizer with error reset 2 unassigned Opto isolated 10 30V input into 3K ohm internal resistor 1 msec minimum pulse width...

Page 76: ...y C 3 AIC 2071 Operating Temperature 13 F to 130 F 25 C to 55 C o o o o Converter Construction Cast aluminum with SBGL corrosion resistant coating Entry Wiring Connections 5 water tight cord grips 4 f...

Page 77: ...0 ohms minimum resistance Prewired CS3 BUS connector IZMSG Hart Transmitter Board Secure ground wire to 300 SMB1 upper left mount with screw Zero Trim 1 Apply loop power to the Hart transmitter board...

Page 78: ...spe b spe D A Trim Loop Test Dac Zero Dac Full DN Pipe Detect Standby Mode Parameter Mode Switch Position Factory Cal Parameters System Cal Parameters Loop Current Cal Meter Setup Display Setup Meter...

Page 79: ......

Page 80: ...FAX 518 922 8997 ANDERSON INSTRUMENT CO LP 400 BRITANNIA RD EAST UNIT 1 MISSISSAUGA ONTARIO L4Z 1X9 CANADA 905 568 1440 FAX 905 568 1652 NEGELE MESSTECHNIK GmbH A Division of Anderson Raiffeisenweg 7...