5.5.
5.5.
5.5.
5.5. Gravity
C
Gravity C
Gravity C
Gravity Compensation
ompensation
ompensation
ompensation F
F
F
Function
unction
unction
unction
If the scale is used at the calibration location, it is not necessary to perform this function.
If there is a difference of gravity acceleration between the installed location and
calibration location it may cause a weighing error. This function specifies the
gravity acceleration and corrects the span error.
Note
The decimal point is not displayed in the function. Example:
9798
= 9.798 m/s
2
When span calibration is executed, the gravity acceleration correction will
be cleared and the two gravity acceleration values will return to the factory
settings.
Step 1 At the
Cal 0
display, press the
MODE
key until
g 5et
is displayed and press the
ENTER
key to enter the gravity compensation function.
If you want to cancel the current procedure, press and hold the
ON/OFF
key and press
the
ESC
key. Then, no parameters are changed and the calibration mode is finished.
Press the
ON/OFF
key to turn the display off after displaying
CanCel
.
Step 2 The parameter is displayed with triangle 1. Enter the gravity acceleration of the
calibration location using the
0
-
9
keys. The parameter xxxx is the gravity
acceleration.
ENTER
key...The key to store the new gravity acceleration and proceed to step 3.
ESC
key .......The key to return to
g 5et
without changing the value.
Step 3 The parameter is displayed with triangle 2. Enter the gravity acceleration of the
installed location using the
0
-
9
keys. The parameter xxxx is the gravity acceleration.
ENTER
key...The key to store the new gravity acceleration and proceed to step 4.
ESC
key .......The key to return to step 2 without changing the value.
Step 4 Now
g
xxxx
is displayed. Press the
CAL
key to store the parameters. The
Caloff
is
displayed. Proceed to step5.
NOTE:
When displaying
Caloff
, press the +/- key while pressing the
ON/OFF
key,
instead of
CAL
key.
Step 5 Press the
ON/OFF
key to turn the display off.
5.5. Gravity Compensation Function