AND TM2441-EX Instruction Manual Download Page 3

 

 

1WMPD4004496 

 

 

The following changes have been made. 

Wrong 

Correct 

#

3 : Example for automated measurement. 

If the start time is specified and 

operation time

 is set 

to 

"OFF"

, when the    AUTO    switch is set 

"ON"

, A-

BPM starts at the preset 

start time

 and continues until 

the    AUTO    switch is set to 

"OFF"

. If the    AUTO   

switch is set to 

"ON"

 again, A-BPM starts at the 

preset 

start time

#

3: If the 

start time

 for automatic measurement is specified 

and the 

operation time

 is set to 

"OFF"

, turning 

"ON"

 

the    AUTO    switch will start automatic blood pressure 
measurement from the 

start time

 set and measurement 

will continue until the    AUTO    switch is turned 

"OFF"

If the    AUTO    switch is turned 

"ON"

 again, automatic 

measurement will continue.

 

If the 

operation time

 is specified and the    AUTO    switch 

is operated during A-BPM, A-BPM is continued for 

operation time

 form this operation. 

If  the 

operation  time

  is  set,  automatic  blood  pressure 

measurement  will  be  performed  from  when  the    AUTO   
switch is initially turned 

"ON"

 until the 

operation time

 has 

passed  even  if  the    AUTO    switch  is  operated  midway 
through measurement.

 

#4 

:

 

Example for automated measurement. 

If the 

start time

 is set to 

"OFF"

 and 

operation time

 is 

specified, when the    AUTO    switch is set to 

"ON"

, A-

BPM is started for the 

operation time

. If the    AUTO   

switch is set to 

"OFF"

 during A-BPM, A-BPM stops. If 

the    AUTO    switch is set to 

"ON"

 again, A-BPM is 

performed for the 

operation time.

 

#4

:  If  the 

start  time

  for  automatic  measurement  is  set  to 

"OFF"

 and the 

operation time

 is set, turning 

"ON"

 the   

AUTO    switch will implement automatic blood pressure 
measurement after initial automatic measurement until 
the  set 

operation  time

  has  passed.  If  the    AUTO   

switch is turned 

"OFF"

 midway through automatic blood 

pressure  measurement,  automatic  measurement  will 
stop.  If  the    AUTO    switch  is  turned 

"ON"

  again, 

automatic measurement will be performed until the set 

operation time

 has passed. 

If the

 start time

 is specified and the    AUTO    switch is 

operated during A-BPM, A-BPM is stopped and is started 
at the preset 

start time 

next day. 

If  the 

start  time

  is  set,  operating  the    AUTO    switch 

midway  through  automatic  blood  pressure  measurement 
will stop automatic measurement and turning it 

"ON"

 again 

will resume measurement. 

 

 

 

The following notes were added to 

"

Example 2 Automatic measurement

"

 

Addition 

NB. Automatic blood pressure measurement will continue even if turned 

"OFF"

 and 

"ON"

 again during 

automatic measurement.

 

 

The following notes were added to 

"

Example 3 Automatic measurement

"

 

Addition 

NB. Automatic blood pressure measurement will not be performed once the 

operation time

 has been 

exceeded even if turned 

"OFF"

 and 

"ON"

 again during automatic measurement. 

 

Instruction Manual Page 

56    8.3.2

. A-BPM Program Examples 

Instruction Manual Page 

53    8.3.1

. A-BPM Items and Parameters 