The following changes have been made.
Wrong
Correct
#
3 : Example for automated measurement.
If the start time is specified and
operation time
is set
to
"OFF"
, when the AUTO switch is set
"ON"
, A-
BPM starts at the preset
start time
and continues until
the AUTO switch is set to
"OFF"
. If the AUTO
switch is set to
"ON"
again, A-BPM starts at the
preset
start time
.
#
3: If the
start time
for automatic measurement is specified
and the
operation time
is set to
"OFF"
, turning
"ON"
the AUTO switch will start automatic blood pressure
measurement from the
start time
set and measurement
will continue until the AUTO switch is turned
"OFF"
.
If the AUTO switch is turned
"ON"
again, automatic
measurement will continue.
If the
operation time
is specified and the AUTO switch
is operated during A-BPM, A-BPM is continued for
operation time
form this operation.
If the
operation time
is set, automatic blood pressure
measurement will be performed from when the AUTO
switch is initially turned
"ON"
until the
operation time
has
passed even if the AUTO switch is operated midway
through measurement.
#4
:
Example for automated measurement.
If the
start time
is set to
"OFF"
and
operation time
is
specified, when the AUTO switch is set to
"ON"
, A-
BPM is started for the
operation time
. If the AUTO
switch is set to
"OFF"
during A-BPM, A-BPM stops. If
the AUTO switch is set to
"ON"
again, A-BPM is
performed for the
operation time.
#4
: If the
start time
for automatic measurement is set to
"OFF"
and the
operation time
is set, turning
"ON"
the
AUTO switch will implement automatic blood pressure
measurement after initial automatic measurement until
the set
operation time
has passed. If the AUTO
switch is turned
"OFF"
midway through automatic blood
pressure measurement, automatic measurement will
stop. If the AUTO switch is turned
"ON"
again,
automatic measurement will be performed until the set
operation time
has passed.
If the
start time
is specified and the AUTO switch is
operated during A-BPM, A-BPM is stopped and is started
at the preset
start time
next day.
If the
start time
is set, operating the AUTO switch
midway through automatic blood pressure measurement
will stop automatic measurement and turning it
"ON"
again
will resume measurement.
The following notes were added to
"
Example 2 Automatic measurement
"
Addition
NB. Automatic blood pressure measurement will continue even if turned
"OFF"
and
"ON"
again during
automatic measurement.
The following notes were added to
"
Example 3 Automatic measurement
"
Addition
NB. Automatic blood pressure measurement will not be performed once the
operation time
has been
exceeded even if turned
"OFF"
and
"ON"
again during automatic measurement.
Instruction Manual Page
56 8.3.2
. A-BPM Program Examples
Instruction Manual Page
53 8.3.1
. A-BPM Items and Parameters