The following changes have been made.
Wrong
Correct
#3
: Example for automated measurement.
Start time :
Stores a time. (0 to 23 hour)
Operation time : Set to
"OFF"
Response :
A-BPM starts the blood pressure measurement
at the preset
Start time
and continues until
A-BPM is suspended.
If the
EVENT
switch is pressing and holding
until the
mark is shown, A-BPM starts at
the preset Start time.
#3
: If the
Start time
for automatic measurement is specified
and the
Operation time
is set to
"OFF"
, long pressing
the EVENT switch will start automatic blood pressure
measurement from the
Start time
set and measurement
will continue until the EVENT switch is long pressed.
If the EVENT switch is long pressed again, automatic
measurement will continue.
Note
If the
Operation time
is set, automatic blood pressure
measurement will be performed from when the EVENT
switch is initially long pressed until the
Operation time
has passed even if the EVENT switch is long pressed
midway through automatic measurement.
#4
:
Example for automated measurement.
Start time :
Set to
"OFF"
Operation time : Stores time to be continued.
(1 to 27 hours)
Response :
A-BPM starts blood pressure measurement
and stops after the
Operation time
.
If the
mark is hidden by pressing and
holding the
EVENT
switch during the
Operation time
, A-BPM stops.
If the
mark is shown by pressing and
holding the
EVENT
switch again, A-BPM is
performed for the
Operation time
.
#4
: If the
Start time
for automatic measurement is set to
"OFF"
and the
Operation time
is set, long pressing the
EVENT switch will implement automatic blood pressure
measurement after initial automatic measurement until
the set
Operation time
has passed. If the EVENT
switch is long pressed midway through automatic blood
pressure measurement, automatic measurement will
stop. If the EVENT switch is long pressed again,
automatic measurement will be performed until the set
Operation time
has passed.
Note
If the
Start time
is set, long pressing the EVENT
switch midway through automatic blood pressure
measurement will stop automatic measurement and long
pressing it again will resume measurement.
The following notes were added to
"
Example 2 Automatic measurement
"
Addition
NB. Automatic blood pressure measurement will continue even if turned
"OFF"
and
"ON"
again during
automatic measurement.
The following notes were added to "Example 3 Automatic measurement"
Addition
NB. Automatic blood pressure measurement will not be performed once the
Operation time
has been
exceeded even if turned
"OFF"
and
"ON"
again during automatic measurement.
Instruction Manual Page
42
8.3.2.
A-BPM Program Examples
Instruction Manual Page
40
8.3.1
. A-BPM Items and Parameters