background image

 

1WMPD4001294C 

 

 

AD-8526

  Ethernet Converter  Instruction Manual

 

 

The AD-8526 Ethernet Converter can connect the RS-232C interface of a weighing instrument to the 

Ethernet (LAN) port of a computer that is not equipped with an RS-232C interface. By using the 

AD-8526, weight data from a weighing instrument can be managed with a computer connected to the 

network. The weighing instrument can also be controlled by the connected computer. 

Unpacking the AD-8526 

 

 

 

 

 

 

 

 

 

 

 

 

Note:  Only the following cable is included if specified when ordering: 

DIN 7-pin to D-Sub 9-pin cable (AX-KO1786-200)

 

If not specified, the following two communication cables are included: 

DIN 7-pin to D-Sub 9-pin cable (AX-KO1786-200), D-Sub 9-pin to D-Sub 9-pin cable (AX-KO2466-200)

 

Specification 

 

1. Specification 

3. RS-232C serial interface 

Operating environment :  -10°C to +40°C 

Connector: 

D-Sub 9-pin male   

AC adapter:  Confirm that the adapter type is correct for 

the local voltage and power receptacle type 

Transmission form:  Asynchronous, bi-directional 

Data format 

Baud rate:  600, 1200, 2400

*

, 4800, 

9600, 19200bps 

Power consumption:  Approx. 11VA (supplied to the AC adapter) 

Data bits:  7

*

 or 8 bits   

Dimensions: 

113(W) x 60(D) x 38(H) mm 

Parity: 

Even

*

, Odd, None   

Net weight: 

Approximately 250g 

Stop bits:  1 bit

*

 or 2 bits   

2. Ethernet interface

 

 

 

Connector: 

RJ45 

 

*

 : Factory settings

 

Protocol: 

TCP/IP 

 

 

Weighing Instrument Cables 

Required AD-8526 Cable by Weighing Instrument 

Weighing instrument 

Interface option 

Communications cable (Length 2 m) 

GX, GF, GX-K, GF-K, GP, GR, HR, 

MC 

None (D-Sub 25-pin, standard accessory) 

AX-KO1710-200 

EK-

i

, EW-

 i

, FC-

 i

, FC-S

 i

, GH, HR-

 i

  None (D-Sub 9-pin, standard accessory) 

AX-KO2466-200 

AD-4212C 

None (D-Sub 9-pin, standard accessory) 

AD4212C standard accessory 

HV-G, HV-WP, HW-G, HW-WP 

None (DIN 7-pin, standard accessory) 

AX-KO1786-200 

FG 

OP-03 (DIN 7-pin) 

AX-KO1786-200 

FG-L, FG-M 

OP-23 (DIN 8-pin) 

AX-KO1786-200 

Please refer to our website for the latest information. 

Connecting the Weighing Instrument to a Computer   

 

Caution 

  Contact your network manager before connecting the AD-8526 to a network. The AD-8526 may 

generate a network error. A&D assumes no responsibility for any errors that may be caused. 

  Affix the AC adapter labels to the AC adapter and the connector end of the cable. 

  Set the IP address and subnet mask to the AD-8526 one at a time.   

The factory default IP address is 172.16.100.2. Do not duplicate an IP address. 

  The IP address cannot be returned to the factory setting.   

We recommend that you write the IP address on the supplied label. 

  Preparation 

Please download each software from A&D website   

https://www.aandd.jp/products/software/software.html

for below.

 

(1) Setting procedure “WinCT-Plus” instruction manual   

(2) IP address setting software “Device Installer” 

(3) Data acquisition software

 

“WinCT-Plus” 

Step 1  Connect the AC adapter to the AD-8526. 
Step 2  Connect the AD-8526 and the computer with a cross 

cable directly or connect them with a hub and straight 

cables. 

Step 3  Input the IP addresses and subnet masks of the AD-8526 

 and computer. For the setting procedure, refer to the 

“WinCT-Plus” instruction manual. 

  Communication error may occur if set incorrectly. 

Step 4  Write the IP address on the IP address label and affix it on 

the AD-8526 (and the weighing instrument

 

if necessary) 

where you can see the address easily. 

Step 5  Install the RsMulti data acquisition software to download the 

“WinCT-Plus” instruction manual.PDF form A&D website. 

 

https://www.aandd.jp/products/software/software.html

 

Step 6  Connect the accessory RS-232C cable between the RS-232C 

interface of the weighing instrument and the AD-8526’s port. 

Data Communication Software 

  The RSMulti data acquisition software can be launched on a Windows PC. Menu: [Start]    [Program]   

[A&D WinCT-Plus]    [RsMulti]. Refer to the Manual PDF file in the A&D WinCT-Plus for details on the 

operation. 

  This software can acquire data from multiple devices connected by LAN or RS-232C. 

  The software can control these devices with commands. 

  The software can acquire data transmitted from devices. Example: When the 

  PRINT 

  key is pressed on the 

balance, data is output and is acquired with the computer. 

  Stored data can be used with Excel. (Install Microsoft Excel before use.) 

Example

   

A sample window of RsMulti

 

  Two balances are connected. 

 

IP address  Subnet mask 

Computer  172.16.100.1 

255.255.0.0 

GX-600 

172.16.100.2 

GP-30K 

172.16.100.3 

 

172.16.100.1 

172.16.100.2  172.16.100.3 

 

GX-600 

GP-30K 

GX-600 

GP-30K 

GX-600 

 

GP-30K 

+0182.252 

+0182.253 

+0182.955 

+0182.964 

+0182.965 

+0186.676 

+0182.251 

+0182.251 

+0182.965 

+0182.265 

+0182.252 

+0182.239 

+019218.1 

+019218.2 

+019218.2 

+019289.8 

+019218.2 

+019218.4 

+019218.0 

+019218.0 

Hub

 

Computer

 

Step 2 

Computer 

Step 2 

RS-232C cable 

Step 6 

Step 6 

AD-8526 

Step1 

Straight cable   

Cross cable 

Weighing 

instrument 

AD-8526 

AC adapter 

 

Note 

Please confirm that 

the AC adapter type 

is correct for your 

local voltage and 

receptacle type. 

IP address label

 (5 labels) 

Write a preset IP address and 

affix it on the weighing instrument.

 

AC adapter labels

 

Affix them to the AC adapter and 

the connector end of the cable. 

IP address protective cover

 

(5 labels)

 

Affix it on the IP address label.

 

RS-232C cable

 

(Length: Approx. 2m. 2 pcs) 

AC adapter

 

AD8526

 

Ethernet converter 

MAC address 

Positions to affix 

AC adapter labels 

(Example) 

Reviews: