AND AD-4430A Instruction Manual Download Page 2

 

 

 

5.  Function Mode 

The function mode stores parameters to control the weighing module. 
The parameters are stored even without power supplied. 
 

5.1.  Summary 

  ............  Calibration function 
  ............  Linearity adjustment function 
  ............  Basics function 
  ............  Hold function 
  ............ Flow rate function 
  ............  Control I/O function 
  ............  Current loop output function 
  ............  Analog 4-20mA output function 

 When only a numerical value is displayed, the decimal point flashes to distinguish 

that from a weighing value.

 

 

5.2.  Operations and Types 

5.2.1.  Select Modes Under Function Mode 

  +        .....Proceeds to function mode from the weighing mode. 

  ...........Selects a type of select mode (3 upper figures).   

  .......Enters a selected mode. 

  ...........Selects an item under the selected mode (2 lower figures).   

  .......Enters the item. 
  .......Stores parameters and returns to weighing mode.

 

 

5.2.2. Change 

Value 

  ...........Moves selected digit. 

  ...........Changes numerical value. 

  .......Activates (store) value and returns to select mode. 
  .......Deactivates value and returns to select mode.

 

 

5.3.  Function Table 

1

  Decimal point depends on 

.

 

2

  When "span calibration using mass (

)" is performed by the input voltage from the 

load cell in the calibration mode, "input voltage at zero (

)", "span calibration input 

voltage (

)" and "weight against span calibration input voltage(

)" change. 

3

  If pressing the 

 key while displaying 

 or 

, the current weighing value 

can be monitored. Press the 

  key again to return function mode.

 

 

5.3.1. Calibration 

Function (

)

 

Item & Function 

Descriptions, Range & 

 

Factory settings

 

 

 

Unit 

 0 :

 No used 

1:

 g 

 2 :

 kg 

3:

 t 

4:

 N 

5:

 kN 

 

Decimal point position 

Decimal point (D.P.) position.   

 0  :

 0 

1 :

 0.0 

2 :

 0.00 

3 :

 0.000 

4 :

 0.0000 

 

Minimum division 

Minimum division of value. (A scale / digit) 

 1  :

 1 

2 :

 2 

3 :

 5 

4 :

 10 

5 :

 20 

6 :

 50 

 

Weighing capacity 

Measurement can be displayed up to +8 digits (8 scales) from 

capacity. 

1

 

1 to   70000    to 99999

 

 

Zero range 

The range that the 

  key (zero key) works. Deviation from the 

calibrated zero point [%]. 

  0 to   2    to 100

 

 

Zero tracking time 

Used with 

 

for zero tracking. 

Scale: 

0.1

 sec. 

 0.0  to 5.0

 

 

Zero tracking band 

Used with 

 

for zero tracking. 

Scale: 

0.1

 digit 

 0.0  to 9.9

 

 

Stability detection time 

Used with 

 

for stability detection. 

Scale: 

0.1

 sec. 

0.0 to   1.0    to 9.9

 

 

Stability detection width 

Used with 

 

for stability detection. 

Scale: 

1

 digit 

0 to   2    to 100

 

 

Tare and zero at unstable 

weight value 

Tare and zero operation when unstable 

 0 : 

Disables both functions. 

 1 : 

Enables both functions. 

 

Tare when the gross 

weight is negative 

Tare when the gross weight is negative.

 

 0 : 

Disables tare. 

 1 : 

Enables tare. 

 

Output when out of 

range and unstable 

 0 : 

Disables output. 

 1 :

 Enables output. 

 

Exceeding negative 

gross weight 

To judge when the negative gross weight is exceeded. 

 1 : 

Gross weight < 

-

99999 

 3 : 

Gross weight < 

-

19 digits 

 2 :

 Gross weight < 

-

Capacity 

 

Exceeding negative net 

weight 

To judge when the negative net weight is exceeded. 

 1 : 

Net weight < 

-

99999 

 2 : 

Net weight < 

-

Capacity 

 

 

Clear the zero value 

Select whether or not to clear the zero value. 

 0 : 

Disable 

 1 :

 Enable 

 

Zero setting when power 

is turned on 

Select whether or not to perform zero setting when power is 

turned on. 

 0 : 

Disable 

 1 :

 Enable 

 
 

 

Input voltage at zero 

Input voltage from a load cell at zero. 

2

 

Scale: 

0.0001

 mV/V 

-7.0000 to  0.0000  to 7.0000

 

 

Span input voltage 

Input voltage from a load cell at span. 

2

 

Scale: 

0.0001

 mV/V 

0.0100 to  3.2000  to 9.9999

 

 

Weight against span 

input voltage 

The calibration weight value corresponding to the input voltage at 

1

 

2

 

1 to   32000  to 99999

 

 

Gravity acceleration of 

the calibration place 

Gravity acceleration of the place where the scale is calibrated. 

Scale: 

0.0001

 m/s

2

 

9.7500 to  9.8000  to 9.8500

 

 

Gravity acceleration of use 

place 

Gravity acceleration of the place where the scale is being used. 

Scale: 

0.0001

 m/s

2

 

9.7500 to  9.8000  to 9.8500

 

 

Suppression of the hold 

function 

 0 : 

Permission. 

 1 :

 Prohibition. 

 ~   

Reserved internally

 

 

 

5.3.2.  Linearity Adjustment Function (

Item & Function 

Descriptions, Range & 

 

Factory settings

 

 

 

Number of input points 

Number of adjustment point. If 0, 1 or 2 is selected, this linearization 

is not used. 

 0  to 5

 

 

Linear-zero 

Input voltage for linear-zero input. 

Scale: 0.0001 mV/V   

-7.0000 to   0.0000    to 7.0000

 

 

Setting value for linear 1 

The setting value of weights for linear 1 input. 

1

 

 

 0  to 99999

 

 

Span at linear 1 

The span voltage between linear-zero and linear 1 input.   

Scale: 0.0001 mV/V   

 0.0000  to 9.9999

 

 

Setting value for linear 2 

The setting value of weights for linear 2 input. 

1

 

 

 0  to 99999

 

 

Span at linear 2 

The span voltage between linear-zero and linear 2 input.   

Scale: 0.0001 mV/V   

 0.0000  to 9.9999

 

 

Setting value for linear 3 

The setting value of weights for linear 3 input. 

1

 

 

 0  to 99999

 

 

Span at linear 3 

The span voltage between linear-zero and linear 3 input.   

Scale: 0.0001 mV/V   

 0.0000  to 9.9999

 

 

Setting value for linear 4 

The setting value of weights for linear 4 input. 

1

 

 

 0  to 99999

 

 

Span at linear 4 

The span voltage between linear-zero and linear 4 input.   

Scale: 0.0001 mV/V   

 0.0000  to 9.9999

 

 

 

5.3.3. Basics 

Function (

)

 

Item & Function 

Descriptions, Range & 

 

Factory settings

 

 

 

Key switch disable 

Each digit corresponds to to a key switch. Only available in weighing 

mode. 

  4 fig.  3 fig.  2 fig.  1 fig. 

0 : 

Permission 

1 :

 Prohibition 

 

 

 

 

 

 

  0 0 0 0    to 1 1 1 1

 

 

    key function 

 0 :

 None 

7:

 Zero clear 

 1 :

  Manual print command 

8:

 Reserved internally 

 2 :

 Hold 

9:

 Reserved internally 

 3 :

  Operation switch 1 

10:

 Reserved internally 

 4 :

  Operation switch 2 

11:

 Reserved internally 

 5 :

 Display exchange 

12:

 mV/V monitor 

 6 :

 Tare clear 

13:

 Digital filter 2 

 

 

14:

 Display output data selected in 

 

 

 

15:

 Display output data selected in 

 

 

Display refresh rate 

 1  :

 20 times/sec. 

2 :

 10 times/sec. 

3 :

 5 times/sec. 

 

 display 

 

 0 :

 None 

 

 

1 :

 Zero tracking in progress 

 

 

2 :

 Alarm 

 

 

3 :

 Display operation switch status as on or off 

 

4 :

 Near

-

zero  

 

5 :

 HI output 

 

 

6 :

 OK output 

 

 

7 :

 LO output 

 

 

Digital filter 1 

Selects a cutoff frequency. 

3

 

 

0 : 

None 

6 :

 20.0 Hz 

12 :

 2.8 Hz 

 

1 :

 100.0 Hz 

7 :

 14.0 Hz 

13 :

 2.0 Hz 

 

2 :

 70.0 Hz 

8 :

 10.0 Hz 

14 :

 1.4 Hz 

 

3 :

 56.0 Hz 

9 :

 7.0 

Hz 

 15 :

 1.0 Hz 

 

4 :

 40.0 Hz 

10 :

 5.6 

Hz 

16 :

 0.7 Hz 

 

5 :

 28.0 Hz 

11 :

 4.0 

Hz 

 

Digital Filter 2 

Selects a cutoff frequency. 

 

0 :

 None 

6 :

 20.0 Hz 

12 :

 2.8 Hz 

18 :

 0.40 Hz 

 

1 :

 100.0 Hz 

7 :

 14.0 Hz 

13 :

 2.0 Hz 

19 :

 0.28 Hz 

 

2 :

 70.0 Hz 

8 :

 10.0 Hz 

14 :

 1.4 Hz 

20 :

 0.20 Hz 

 

3 :

 56.0 Hz 

9 :

 7.0 

Hz 

 15 :

 1.0 Hz 

21 :

 0.14 Hz 

 

4 :

 40.0 Hz 

10 :

 5.6 

Hz 

16 :

 0.7 Hz 

22 :

 0.10 Hz 

 

5 :

 28.0 Hz 

11 :

 4.0 

Hz 

17 :

 0.56 Hz 

23 :

 0.07 Hz 

 

Hold function 

 1 :

 Normal hold 

2 :

 Peak hold 

3 :

 Averaging hold 

 

 

5.3.7.  Current loop Output Function (

Item & Function 

Descriptions, Range & 

 

Factory settings

 

 

 

Serial data 

 1 :

 Displayed weighing value   

3 :

 Net   

5 :

 Gross / Net / Tare 

 2 :

 Gross 

 

4 :

 Tare 

 

Communication mode 

 1 :

 Stream 

3 :

 Manual print 

 2 :

 Automatic print 

 

Baud rate 

 1 :

 600 bps 

 2 :

 2400 bps 

 

 

5.3.8. Analog 

4-20mA

 

Output Function (

)

 

Item & Function 

Descriptions, Range & 

 

Factory settings

 

 

 

Output data

 

 1 :

 Displayed weighing value (Digital filter 1)

 

2 :

 Gross (Digital filter 1)

 

3 :

 Net (Digital filter 1)

 

4 :

 Displayed weighing value (Digital filter 2)

 

5 :

 Gross (Digital filter 2) 

6 :

 Net (Digital filter 2)

 

7 :

 Flow rate 1

 

8 :

 Flow rate 2

 

9 :

 Flow rate 1 or Flow rate 2 (Select in control input)

 

 

Mass/flow rate 

at 4mA output 

Select mass/flow rate by setting output data (

Decimal point position linkage: 
 Mass 

 

 Flow rate 

 (setting magnification of flow rate) + 

 

 

-

99999 to    0   to 99999 

 

Mass/flow rate 

at 20mA output 

Select mass/flow rate by set output data (

Decimal point position linkage: 
 Mass 

 

 Flow rate 

 

(setting magnification of flow rate) + 

 

 

-

99999 to  70000  to 99999

 

 

Flow rate unit 

 1 :

 Seconds 

2 :

 Minutes 

3 :

 Hours 

IOUT1

 

 

Flow rate setting 

magnification 

(times)

 

 

 1 :

 

 1 

2 :

 

 10 

3 :

 

 100 

4 :

 

 1000 

5 :

 

 10000

 

 

Output data 

 1 :

 Displayed weighing value (Digital filter 1) 

2 :

 Gross (Digital filter 1) 

3 :

 Net (Digital filter 1) 

4 :

 Displayed weighing value (Digital filter 2) 

5 :

 Gross (Digital filter 2) 

6 :

 Net (Digital filter 2) 

7 :

 Flow rate 1 

8 :

 Flow rate 2 

9 :

 Flow rate 1 or Flow rate 2 (assign to input terminals) 

 

Mass/flow rate 

at 4mA output 

Select mass/flow rate by set output data (

Decimal point position linkage: 
 Mass 

 

 Flow rate 

 (setting magnification of flow rate) + 

 

-

99999 to    0    to 99999 

 

Mass/flow rate 

at 20mA output 

Select mass/flow rate by set output data (

Decimal point position linkage: 
 Mass 

 

 Flow rate 

 (setting magnification of flow rate) + 

 

 

-

99999 to  70000  to 99999 

 

Flow rate unit 

 1 :

 Seconds 

2 :

 Minutes 

3 :

 Hours 

IOUT2

 

      

Flow rate setting 

magnification 

(times) 

 1 :

 

 1 

2 :

 

 10 

3 :

 

 100 

4 :

 

 1000 

5 :

 

 10000

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

 

Near

-

zero

 

1

 

-99999 to  10  to 99999

 

 

 

Comparison mass at 

near-zero 

 1 :

 Gross weight 

2 :

 Net weight 

 

Upper limit value 

1

 

-99999 to  10  to 99999

 

 

Lower limit value 

-99999 to  -10  to 99999

 

 

Comparison mass of 

upper and lower limit 

 1 :

 Gross weight 

2 :

 Net weight 

 

5.3.4. Hold 

Function 

(

Item & Function 

Descriptions, Range & 

 

Factory settings

 

 

 

Average time 

Time to calculate the average. 

0.00

 is not averaged. 

Scale: 0.01 sec.   

 0.00  to 9.99

 

 

Start wait time 

Waiting time to commence a holding or averaging. 

Scale: 0.01 sec.   

 0.00  to 9.99

 

 

Condition of 

automatic start 

The condition to commence a holding or averaging. 

 0 :

 Not used 

2 :

 Above the near-zero 

 

1 :

 Above the near-zero, and stable 

 

Release using control 

input 

Release when control input of the hold terminal is falling. 

 0 :

 Do not release 

 1 :

 Release 

 

Release time 

Release after a set amount of time has passed. 

0.00 

is not averaged. 

Scale: 0.01 sec.   

 0.00  to 9.99

 

 

Release using 

fluctuation range 

Release when fluctuation from the holding value exceeds a set value.   

0 :

 Continue 

 0  to 99999

 

 

Release at near-zero 

Release when the weighing value is in the near-zero range. 

 0 :

 Do not release 

1 :

 Release 

 

5.3.5.  Flow Rate Function (

)

 

Item & Function 

Descriptions, Range & 

 

Factory settings

 

 

 

Filter of flow rate 1

 

 

Filter of flow rate 2

 

1 :

Digital filter 1

 

12 :

Digital filter 2

 

 

Damping time of flow rate 1

 

 

Damping time of flow rate 2

 

Suppress shaking of flow rate. 
The higher value setting, the less shaking.

 

Scale: 1 sec.

 

1 to    5    to 1000

 

 

 

5.3.6.  Control I/O Function (

)

 

Item & Function 

Descriptions, Range & 

 

Factory settings

 

 

 

IN 1 

0 to  7 to 30 

 

IN 2 

0 to  8  to 30 

  IN 3 

 0  to 30 

 

IN 4 

 0  to 30 

  IN 5 

 0  to 30 

IN

 

  IN 6 

 0 :

 Not used 

 

22 :

 Zero clear 

  1 to 6 :

 Reserved internally 

23 :

 Tare clear 

 7 :

 Zero 

 

24 :

 Operation same as a    F    key 

 8 :

 Tare 

 

25 :

 Prohibit update of flow rate 1 

 9 :

 Hold 

 

26 :

 Prohibit update of flow rate 2 

10 :

 

Gross 

/ Net exchange 

27 :

 Initialize flow rate 1 

11 :

 Diagnose 

 

28 :

 Initialize flow rate 2 

12 :

 Print command     

29 :

 Specify flow rate in 

 

 

13 to

 

21 : 

Reserved internally

 30 :

 Specify flow rate in 

 

 

 

 

 

Select flow rate 1 or 2 

 0  to 30 

  OUT 1 

0 to 18  to 36 

  OUT 2 

0 to  9  to 36 

  OUT 3 

 0  to 36 

  OUT 4 

 0  to 36 

  OUT 5 

 0  to 36 

  OUT 6 

 0  to 36 

  OUT 7 

 0  to 36 

  OUT 8

 

 0 :

 Not used 

 

 

  1 to 8 : 

Reserved internally   

 9 :

 Stability 

 

 

10 :

 Over capacity 

 

30 :

 In weighing (ON) 

11 :

 Net display 

 

31 :

 In weighing (1 Hz) 

12 :

 During tare 

32 :

 In weighing (50 Hz) 

13 :

 Hold 

33 :

 Alarm 

14 :

 Hold busy 

34 :

 Output operation switch is   

15 :

 HI output 

  on or off 

16 :

 OK output 

35 : 

Approximate flow rate value  

17 :

 LO output 

  of flow rate 1 

18 :

 Near-zero 

36 :

 Approximate flow rate value   

19 to 29 :

 Reserved internally   of flow rate 2 

 0  to 36 

  OUT 1 
  OUT 2 

  OUT 3 

  OUT 4 

  OUT 5 
  OUT 6 

  OUT 7 

OUT
 

  OUT 8

 

 1 :

 Inverting output 

 2 :

 Non inverting output

 

 

 

 

 

 

-

5

-

6

-

7

-

8