54 of 220
Anatel A643a Setup - Anatel A643a
R
PS
-
June
20
07
-
Ed
it
io
n
12
Operator Manual
Anatel
5.3
Password Security
Note:
Defaults —
Identifier: Super
Supervisor: 00000 (None)
Identifier: Opera
Operator: 00000 (None)
The Anatel A643a’s operational modes (Setup, Manual and Print) can be password protected
to prevent unauthorized users from accessing any of the instrument’s configuration or control
parameters, as well as its printout functions. The Anatel A643a allows two levels of password
security: a Supervisor who has access to all functions and an Operator who can alter
everything but the password entries and the
Log Setup Menu
(see
)
which contains the instrument’s audit trail.
The optional passwords are equipped with a required
Identifier
to trace accountability. Any
parameter changes are recorded in the Analyzer’s internal log for later retrieval and are
automatically output to the printer (assuming one is attached). These records include the time,
the parameter and the date of the change in addition to identifying the person authorized to
make the modification.
By default, no passwords are defined (indicated by settings of
00000
), thereby allowing
complete access to all personnel. A password, once enabled, is constrained by a 90-day limit.
In other words, active passwords must be changed every three months. An alarm code is
issued three times a day—at 12:00 AM, 8:00 AM and 4:00 PM—when the assigned limit has
expired. This code will continue to be displayed (and printed) until the Supervisor changes the
current password settings. To further enhance this security feature, the same passwords
cannot be used in two successive 90-day periods.
Note:
Settings of 00000 for the Supervisor and the Operator disable this security feature.
To establish a supervisory Password and Identifier:
1)
With the desired Analyzer selected in any View, press the
Setup
Key
to display its
submenu.
2)
Use the
Up
and
Down Keys
to specify
Sensor Setup
.
Fig 5-4 :
Set Password Menu
Summary of Contents for A643
Page 1: ...Operator Manual ANATEL A643a TOC ANALYZER...
Page 2: ......
Page 18: ...18 of 220 Introduction Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 80: ...80 of 220 Anatel A643a Setup Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 110: ...110 of 220 Anatel A643a Alarms Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 120: ...120 of 220 Printer Output Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 138: ...138 of 220 Analog Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 144: ...144 of 220 Digital Inputs Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 154: ...154 of 220 Serial Communications Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 188: ...188 of 220 Troubleshooting Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 190: ...190 of 220 Service Procedures Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 212: ...212 of 220 Glossary Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 219: ...Anatel A643a User Notes 219 of 220 RPS June 2007 Edition 12 Anatel Operator Manual User Notes...
Page 221: ......
Page 222: ......