Once all eight of the switches have been set, a printout prompts whether to “Continue”
to repeat the procedure for
DIP Switch 2
and then
DIP Switch 3
, or “Write” the current
settings to memory. Once saved, a printout is generated and the printer returns to
normal operation.
CAUTION:
Do not turn the printer off while it is saving settings—wait until Dip SW setting complete!
is printed before removing power.
8.3
Software Setup
Hard copies of Analyzer data can be produced on demand and other print-related operations
can be performed. The available print options include the following functions:
•
The ability to print and examine the Analyzer’s internal log.
•
The capacity to print comprehensive Analyzer setup and factory configuration
parameters.
8.3.1
Internal Log
Note:
Default: Data & Audit
Each Anatel A643a Analyzer’s onboard log protects against data loss by chronologically
recording any events. An
Audit Only
selection in the
Log Setup Menu
records power cycles,
alarm codes and parameter changes. The default setting of
Data & Audit
also stores analysis
results and Purge Mode data. In either case, all events are time-stamped and the log is only
updated when information is sent to the printer. Any parameter changes have an additional line
containing the date and identifying the person who made the instrument change.
The log also tracks the Analyzer’s available memory. The amount of data stored in the log
depends on the
Log Setup
as well as the instrument’s
TOC Print
) and
Purge Print
) settings. For example, with
Data
& Audit
selected and sampling at one-minute intervals in the Purge Mode, the log will retain
approximately two and a half days worth of information before it begins writing over the earliest
data recorded; sampling every six minutes would hold about 16 days of data. With
Audit Only
selected and few changes made to the instrument, the log will store information for an indefinite
period.
Summary of Contents for A643
Page 1: ...Operator Manual ANATEL A643a TOC ANALYZER...
Page 2: ......
Page 18: ...18 of 220 Introduction Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 80: ...80 of 220 Anatel A643a Setup Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 110: ...110 of 220 Anatel A643a Alarms Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 120: ...120 of 220 Printer Output Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 138: ...138 of 220 Analog Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 144: ...144 of 220 Digital Inputs Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 154: ...154 of 220 Serial Communications Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 188: ...188 of 220 Troubleshooting Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 190: ...190 of 220 Service Procedures Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 212: ...212 of 220 Glossary Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 219: ...Anatel A643a User Notes 219 of 220 RPS June 2007 Edition 12 Anatel Operator Manual User Notes...
Page 221: ......
Page 222: ......