Transport and storage
58
Edition 06/2015
InnuPure C16
Switch off the device and disconnect the mains connection. Remove the HID-Pro
320 from the device.
To prevent damage to health, the following must be observed when moving the
device in the laboratory (lifting and carrying):
–
For reasons of safety, two persons are required to transport the device and
must position themselves on both sides of the equipment.
–
Because the device does not have carrying handles, grip the device firmly with
both hands at the lower device end, but not in the door area. Make sure the
doors are closed before lifting the device simultaneously.
–
Observe the guide values and adhere to the legally mandated limits for lifting
and carrying without auxiliary means!
For the setup at the new location observe the notes in the section "Site require-
ments" p. 22.
9.2
Storage
ATTENTION
Environmental influences and condensate formation can destroy individual components of
the device!
The device must only be stored in air-conditioned rooms. The atmosphere must be low in
dust and free from aggressive vapors.
If the device is not installed immediately after delivery or not required for prolonged periods,
it should be stored in its original packaging. A suitable desiccant should be added to the
equipment to prevent damage from moisture.
The following storage conditions must be met:
Temperature range:
-10 – 55°C
Max. humidity:
max. 80% (use desiccant)