Analox 50 – Carbon Dioxide Analyser – User Manual 

 
 

 

Document Ref:  A50-811-19 – September 2014 

 

 

   

 

      Page 14 

 

 

 

 

 

 

 

 

 

4.2.2.  To connect a “Quick Connect” Remote Alarm Repeater 

 

1)  Disconnect the power supply from the Ax 50

TM

2)  Insert    the connector on the end of the wire into either socket on 

the bottom of the Remote Alarm Repeater. 

3)  Restore power to the Ax 50

TM

. Press the switch on the Remote 

Alarm Repeater once, and ensure that the four indicators flash. 
Note that in the presence of a genuine alarm, this test feature is 
disabled. 

4.2.3.  To connect a “Strobe LED” Remote Alarm Repeater 

 

1)  Disconnect the power supply from the Ax 50

TM

2)  Open the Remote Alarm Repeater unit by removing the 4 screws 

in the front of the case and carefully pulling the case apart. The 
connecting wires from the Ax 50

TM

 pass through one of the cable 

glands on the Remote Alarm Repeater.   

3)  Terminate the wires in accordance with the table below 

 

 

4) For units 

NOT

 fitted with internal relay(s): Connect the free wire 

(Brown or Orange) to A50-231 main PCB at JP13 terminal 5. 
For units fitted with internal relay(s): Connect the free wire (Brown 
or Orange) to A50-231 main PCB at IC10 terminal 2 and move the 
wire from terminal 4 (Green/white or green) to terminal 2 of ST1. 

5)  Replace the lid of the Remote Alarm Repeater, insert and tighten 

the 4 screws, and mount it in the desired position using the 2 
fixing lugs. 

REPEATER CABLE COLOURS 

Ax 50 

Type 1 

Repeater 

Term 

Type 2 (Cat 5) 

ST1 

Terminal 

No 

  

                                

  

                                

*See 

Note 4 

  

                                

  

                                

  

  

  

 

 

 

 

 

 

 

 

 

 

 

 

  

  

                                

  

                                

  

                                

Summary of Contents for Ax 50

Page 1: ...kshire TS9 5PT UK T 44 0 1642 711400 F 44 0 1642 713900 W www analox net E info analox net This support line is closed on UK public holidays Ax 50 Carbon Dioxide Analyser WARNING Please read the safet...

Page 2: ......

Page 3: ...Analox 50 Carbon Dioxide Analyser User Manual Document Ref A50 811 19 September 2014 Page 2 How to fit an Ax 50TM 2 450mm 1...

Page 4: ...Analox 50 Carbon Dioxide Analyser User Manual Document Ref A50 811 19 September 2014 Page 3 3 4...

Page 5: ...Analox 50 Carbon Dioxide Analyser User Manual Document Ref A50 811 19 September 2014 Page 4 5 6...

Page 6: ...Analox 50 Carbon Dioxide Analyser User Manual Document Ref A50 811 19 September 2014 Page 5 7 8...

Page 7: ...Analox 50 Carbon Dioxide Analyser User Manual Document Ref A50 811 19 September 2014 Page 6 Ax 50TM Accessories b h g e c i a...

Page 8: ...fects of Carbon Dioxide 1 000ppm 0 1 5 000ppm 0 5 10 000ppm 1 15 000ppm 1 5 20 000ppm 2 30 000ppm 3 40 000 50 000ppm 4 5 50 000 100 000ppm 5 10 100 000 1 000 000ppm 10 100 For a more detailed outline...

Page 9: ...onnect a Hard Wired Remote Alarm Repeater 13 4 2 2 To connect a Quick Connect Remote Alarm Repeater 14 4 2 3 To connect a Strobe LED Remote Alarm Repeater 14 4 3 Alarm Relay Output Models 15 4 4 Relay...

Page 10: ...Repeater is pre wired to the Ax 50TM sensor unit c If you need to disconnect the cable for ease of installation DISCONNECT FROM THE REMOTE ALARM REPEATER END d The system MUST NOT be switched on unti...

Page 11: ...e The Remote Alarm Repeater mimics the status indicators on the main Ax 50TM It also provides a push button which operates in the same manner as the Mode switch on the Ax 50TM If you have more than on...

Page 12: ...ower failure some examples are listed below Instrument Type Battery Back up Endurance Standard Ax 50TM with 1 Remote Alarm 14 Hours Ax 50TM including 4 20mA Output 1 Relay Output and 2 Remote Alarms 1...

Page 13: ...l must be placed by the local disconnection point for the Ax 50TM stating REMOVE PLUG TO DISCONNECT Ax 50TM The Ax 50TM should be mounted onto a wall using the mounting lugs on the unit Rawl plugs and...

Page 14: ...g the 4 screws in the front of the case and carefully pulling the case apart The connecting wires from the Ax 50TM pass through one of the cable glands on the Remote Alarm Repeater 3 Terminate the wir...

Page 15: ...ter unit by removing the 4 screws in the front of the case and carefully pulling the case apart The connecting wires from the Ax 50TM pass through one of the cable glands on the Remote Alarm Repeater...

Page 16: ...e Volt Free single pole Changeover and are rated 50VAC DC 1 Amps max External relays A pre wired junction box is supplied with the Ax 50TM main sensor unit There is no need to open the main sensor uni...

Page 17: ...gland must be used Ensure that the gland is properly tightened Test that the cable is adequately gripped by the cable gland Ensure that the cable is suitable for purpose the load is within the limits...

Page 18: ...a 4 to 20mA output A 2m cable is factory fitted to the instrument The 4 to 20mA output current is generated by the instrument The customer system should be connected to this cable as follows and as pe...

Page 19: ...gas molecules not the percentage in the atmosphere therefore at altitudes above 900 metres 3000 feet alarms will operate below the factory calibration point Please refer to our website www analox net...

Page 20: ...ioxide continues to rise above the third alarm level then the red third Alarm indicator will also begin to flash and the pace of the Horn will increase This condition will be repeated on any Remote Al...

Page 21: ...procedure As long as the green light is flashing the instrument is working If there are no indicator lamps lit on the Ax 50TM check that power is connected and that the fuses are OK 1 If the OK indica...

Page 22: ...l Display Optional 4 to 20mA Output or 0 1V Gas Concentration Output Max load 150 Alarms 2 3 x Carbon Dioxide Visual Indicators 1 x System Fault Indicator 1 x Status Indicator 1 x Optional Oxygen Visu...

Page 23: ...e but in our discretion we may elect instead to provide to you which ever is the lesser of the cost of replacement or a refund of net purchase price paid as per our Invoice on initial purchase from us...

Page 24: ...ument Ref A50 811 19 September 2014 Page 23 9 DISPOSAL According to WEEE regulation this electronic product can not be placed in household waste bins Please check local regulations for information on...

Page 25: ...Alarm Repeater Battery Backup and Relay Terminal Junction Box all have ingress protection to IP65 low pressure water jets from all directions and totally protected from dust in accordance with EN 605...

Page 26: ...Analox 50 Carbon Dioxide Analyser User Manual Document Ref A50 811 19 September 2014 Page 25 11 DECLARATION OF CONFORMITY...

Page 27: ......

Reviews: