background image

 

ANALOX 1371 MK 2 SURVEYOR CARBON DIOXIDE MONITOR

 

 
 
 

 

 
1371-810-03 User Manual, August 2007 

1317

 

MAINTENANCE

 

      

6.1  

GENERAL 

 

 

Routine maintenance should be carried out at six monthly intervals to  ensure continued reliable 
operation.  The procedure detailed below should be followed and should not take more than about 
20 minutes.  It  is strongly recommended that the procedures be carried out by a technician or 
engineer, familiar with Electronic Instrumentation. 

 

Scottish Anglo E.P. Ltd offer both a full calibration service, and training service for customers' 
maintenance personnel. Full details are available from Scottish Anglo E.P. Ltd. Customer services 
Department, who may be contacted at the Address / Telephone / Fax shown on the  front page of 
this manual.   

      

6.2  

CHANGING THE FILTER

 

 

Disable the power supply to the Instrument and open the front cover. Remove the four corner 
screws from the front panel  and  carefully  ease the front panel away from  the  Instrument case.  
The  panel is attached to the main circuit board by some  wires but it can be temporarily 'parked'  in 
the uppermost slots on the left  and right sides of the case.  Locate  the  plastic filter unit fitted in 
the Gas Inlet Pipe. Carefully remove the plastic tubing from the ends of  the and insert the new 
filter in the piping.  Replace the front panel, ensuring that the wires  to  the front panel are not 
trapped between the panel  and  the case.  The Instrument may now have its power supply restored 
and after initial warm-up will be ready for operation.     

 

Replacement Filters are available from Scottish Anglo E.P.  Ltd. and should be ordered as Part 
Number 1000-1401 

 

A full Calibration procedure,  as detailed in Section 5 should  now be carried out. 

      

NOTE  :   If  the  Instrument  is  being used in a  particularly   dusty atmosphere,  the  filter  may  
require changing more frequently. 

Summary of Contents for 1371 MK2

Page 1: ...R 1371 810 03 User Manual August 2007 117 ANALOX 1371 MK2 SURVEYOR CARBON DIOXIDE MONITOR INSTRUCTION MANUAL Analox Sensor Techn 15 Ellerbeck Court Stokesley Business Park STOKESLEY Cleveland TS9 5PT...

Page 2: ...ONING 8 3 1 INSTALLATION 8 3 2 CONNECTIONS 9 3 3 COMMISSIONING 9 4 OPERATION 10 4 1 ALARM SETPOINTS 10 5 CALIBRATION 11 5 1 GENERAL COMMENTS 11 5 2 CALIBRATION PROCEDURE 11 6 MAINTENANCE 14 6 1 GENERA...

Page 3: ...o be correct at the time of publication The purchasers attention is drawn to our terms and conditions of sale a copy of which is available on request Further technical assistance may be obtained by co...

Page 4: ...ANALOX 1371 MK 2 SURVEYOR CARBON DIOXIDE MONITOR 1371 810 03 User Manual August 2007 417...

Page 5: ...ndependent user adjustable alarm circuits are incorporated the setpoints of which are accessible on the front panel and adjustable over the full range of the Instrument Both audible and visual indicat...

Page 6: ...larm Ranges Adjustable over full range of Measurement 2 2 OPERATIONAL Ambient Temperature 5 to 50 C Wider operating Temperature Units are available on request Storage Temperature 20 to 70 C Analyser T...

Page 7: ...ensions 296mm W 256mm H 118mm D Weight 3 2 kg Gas Connections Suitable for 6mmOD 4mmID plastic Pipe NOTE DISPLAY RESOLUTION ACCURACY The Digital displays used in this series of instruments are of the...

Page 8: ...e two screws from the Terminal compartment cover situated along the lower edge of the Instrument Remove the cover taking care not to strain the connecting wires to the audible warning buzzer These wir...

Page 9: ...the Terminal compartment 3 2 3 On DC Supply versions a 12V DC supply should be connected to the terminals marked 12V and 3 2 4 Remote gas sampling may be carried out by connecting up to about 30 metr...

Page 10: ...tate The relayand LED will remain in this condition until the measured concentration falls below the alarm setpoint at which time the relay will revert to its normal state and the LED will be turned o...

Page 11: ...ed this is quite normal Connect the delivery tube to the small Barbed connector on the flowmeter and then to the GAS IN port on the Instrument Adjust the rotary valve on the flowmeter until the indica...

Page 12: ...045 5 2 5 If during the Calibration process the desired readings cannot be achieved then it is possible that the Infra Red Transducer module requires service To check that the Transducer module is ope...

Page 13: ...nstrument and open the front cover Remove the four corner screws from the front panel and carefully ease the front panel away from the Instrument case The panel is attached to the main circuit board b...

Page 14: ...ange of measurement and the gases can be identified by use of narrow band interference filters tuned to an appropriate characteristic absorption wavelength The family of Transducers used in the ANALOX...

Page 15: ...rn to us would be repaired or replaced without cost or charge but in our discretion we may elect instead to provide to you which ever is the lesser of the cost of replacement or a refund of net purcha...

Page 16: ...ems must be properly and sufficiently packed 8 2 REPAIR Apart from routine Calibration and filter changes there are no user serviceable parts inside the Instrument In the event of any suspected malfun...

Page 17: ...f the housing 2 The Transducer is located in the lower right corner of the housing It is a metal box measuring approximately 115mm x 65mm x 30mm 3 Remove the 6mm OD plastic tubes from the gas flow ada...

Page 18: ...Housing If obviously not operating contact Agent or SAEP Ltd Check any external tubing to the Instrument for blockage kinks No Display response to Calibration Gas Check Calibration Gas Cylinder is not...

Page 19: ...0 10 FS SA7 D03 10P 20 FS SA7 D03 20P 100 FS SA7 D03 100P 10 2 CALIBRATION SPARES Gases 20 Litre Cylinder Disposable Test Gas Pure Nitrogen SA7 L20 01 0 5 CO2 in Air SA7 L20 130 1 5 CO2 in Air SA7 L20...

Reviews: