Analog Technologies ATTECA200WC Quick Start Manual Download Page 3

 

 

 

 

1161 Ringwood Ct, #110, San Jose, CA 95131, U. S. A. Tel.: (408) 748-9100, Fax: (408) 770-9187                           www.analogtechnologies.com 

Copyrights 2000-2021, Analog Technologies, Inc. All Rights Reserved. Updated on 3/18/2021   

 

  Email: [email protected]/[email protected]             3     

Analog Technologies 

ATTECA200WC

 

 

TEC Air Conditioner

 

MENU CODE SELECTION 

 

Figure 5. Physical Photo of the Control Panel 

Table 1. Menu code description 

Symbols 

Description 

Setting Range 

Default Settings 

Unit 

F0

 

Hysteresis for heating 

0

 

~

 

15 

°C

 

F1

 

Hysteresis for cooling 

0

 

~

 

15 

°C 

F2 

Temperature limit setting

 

−9

 

~

 

F3 

15 

°C 

F3

 

Setting upper temperature limit

 

F2

 

~

 

99 

30 

°C 

F4 

Temperature correction 

−7

 

~

 

+7 

°C 

F5 

Delay start 

0

 

~

 

minute 

F6 

Low temperature alarm output 

−9

 

~

 

F6 

−9 

°C

 

F7 

High temperature alarm output

 

F7

 

~

 

99 

70 

°C

 

Press 

Rst

 to set the TEC air cooler. Press ▲ or ▼ to 

increase or decrease the parameter value. 

Set

 

light

 is on. 

Press 

Set

 to exit the setting. 

Constant Temperature Control 

1. When the measured temperature ≥ F3, start the cooling 
output; when the measured temperature ≤ F3

 

 

F1, shut 

down the output. 

2. When the measured temperature value < F2, start the 
heating output; when the measured temperature ≥ F2

 

+

 

F0, 

shut down the output. 

Constant Temperature Setting 

Press the 

Rst

 to enter the menu, and press 

 or ▼ until 

"F2" or "F3" code is displayed. Press 

Set

 to display the 

temperature setting value, and then press the 

 or ▼to 

adjust the temperature value. 

"F2" indicates the lowest temperature in the temperature 
range, "F3" is the highest temperature in the temperature 
range. 

 

 

Hysteresis 

Hysteresis is the difference between the temperature at 
which the controller switches off and the temperature at 
which it switches on again. The minimum difference is 0°C, 
and the maximum 15°C.

 

Press the 

Rst

 to enter the menu,

 

and then press 

 or 

 

until "F0" or "F1" code is displayed. Press 

Rst

 to display 

the setting value of hysteresis, and then press the 

 or ▼ 

to adjust the value. "F0" refers to the hysteresis for heating, 
and "F1" is the hysteresis for cooling. 

Temperature Calibration 

When the measured temperature is different from standard 
temperature, use this temperature calibration function to 
make these two temperatures consistent. Calibrated 
temperature = temperature before calib calibrated 
value (positive, negative or 0). 

Press the 

Rst

 to enter the menu,

 

and then press 

 or 

 

until "F4" is displayed. Press 

Rst

 to display the 

temperature calibration value, and then press the 

 or ▼ 

to adjust the value.

 

 