Analog Devices │
5
MAX22005PMB#
Evaluates: MAX22005
11) Voltages can be measured on the four input chan
-
nels independently. This example measures all four at
once with the same voltage. Select all four Channels
in
Measurement Configuration
by checking each
box. In the same section, select the
Physical Values
button. In the
Results
box on the right side,
Voltage
should be selected (the button switch is on the left
side). See
12) Connect a voltage source from the CH1 (Pin 1), CH2
(Pin 4), CH3 (Pin 7), and CH4 (Pin 10) pins to GND.
Remember to connect Pin 2 to Pin 3 (GND), Pin 5 to
Pin 6 (GND), Pin 8 to Pin 9 (GND), and Pin 11 to Pin
12 (GND) for proper operation with the Munich GUI.
Set the voltage source to +10V and activate it.
13) In the
Sample
box, set the
Sample Rate
and
Sam-
ple Count
.
Figure 5. Example Voltage Measurement Settings and Expected Results
Figure 6. Example Sample Settings