UG-1274
Rev. 0 | Page 5 of 10
ACQUIRING DATA
SELECTING THE PROPER VIEW
The EVAL-ADPD188BIZ-SK is intended for smoke and aerosol
detection applications. Select the
Smoke Device
data view (see
Figure 10) to open a window that allows the user to run the
device and to collect data (see Figure 19).
16756-
010
Figure 10. Selecting the
Smoke Device
Data View
LOAD CONFIGURATION
In the upper right corner of the data view window, click the
ADPD
Config
button (see Figure 11) to open the
Configuration
window.
16
756-
0
11
Figure 11. Selecting
ADPD Config
to the Open the
Configuration
Window
In the
Configuration
window, click
Load DCFG
to select a
configuration file, as shown in Figure 12. For smoke and aerosol
measurements, select the
ADPD188BI_SK.dcfg
configuration
file from the file dialog box.
16
75
6-
0
12
Figure 12. Selecting
Load DCFG
to Select the Configuration File
OPTIMIZING AND RUNNING THE
Using the
ADPD188BI_SK.dcfg
configuration file, the
is set up with an infrared (IR) LED firing in
Time Slot A and the blue LED firing in Time Slot B. The
initial configuration is as shown in Figure 18. Before running the
, a few steps must be taken to optimize
the data collection view and provide valid blue/IR ratios.
1.
In the data view (see Figure 19), click the
Op Select
button
to open the view shown in Figure 13. Select
SB_Sum
,
SA_Sum
, and the
/
operation, as shown in Figure 13, to
configure the blue/IR ratio in the data view.
16756-
013
Figure 13.
Op Select
View
2.
Before running the WaveTool, make sure the operating
mode shown in the
Set SlotMode
window of the data view
is set to
ADPDDrv_SUM_32
for both Time Slot A and
Time Slot B, as shown in Figure 14.
1675
6-
014
Figure 14. Selecting
ADPDDrv_SUM_32