Amvex VR-CIU2-F2D Operating And Maintenance Manual Download Page 5

5

Procedures Prior to Use List:

   

     

The following tests are recommended 

prior to use on each patient

                             If the Vacuum Regulator does not pass one or more of the following tests, 
                             it should be evaluated, repaired and/or replaced by a qualifi ed  individual.    

The following tests must be done with a minimum supply vacuum of -53 kPa(-400 mmHg):

1.  Move the selector switch to the “OFF” position. Turn the regulator knob one complete    
     turn in the clockwise direction. Kink the vacuum tubing to block the outlet. There should 
     be no movement of the gauge needle (or in case of a digital gauge, no change in display).

2.  Move the selector switch to the “REG” position. Turn the regulator knob fully in the 
     counter-clockwise direction. Kink the vacuum tubing; again, there should be no 
     movement of the gauge needle (or in case of a digital gauge, no change in display).

3.  Kink vacuum tubing.

Regulator Setting:

Standard: 

                    Increase the vacuum to -12 kPa (-90 mmHg)

Pediatric & Neonatal:

 Increase the vacuum to -5 kPa (-40 mmHg)

4.  Open and close the kinked vacuum tubing slowly to reach various vacuum rates. Ensure that 
     the level of vacuum maintains consistently when the vacuum tubing is kinked.

5.  Move the selector switch to “INT”.

6.  Kink vacuum tubing.

7.  Timing cycles are approximately 16 seconds on and 8 seconds off.

NOTE:

                The intermittent unit starts in the off cycle. 

8.  Decrease the vacuum level to zero and move the selector switch to the “OFF” position.

Pediatric & Neonatal:

9.  In the “REG” position, kink the vacuum tubing and turn the regulator control knob fully in the 
     clockwise direction to ensure that the vacuum level does not go over -21 kPa (-160 mmHg) for 
     Pediatric and -13 kPa (-100mmHg) for Neonatal.

NOTE:                  

This feature is only present in the Pediatric and Neonatal models.

10. Decrease the vacuum level to zero and move the selector switch to the “OFF” position.

    

      Always verify vacuum setting prior to performing any procedure. Vacuum  

    

      levels set in the “REG” mode will remain the same when switched to the  

    

     “INT”  mode; and vacuum levels set in the “INT” mode will remain the same  

    

     when switched to the “REG” mode. 

CAUTION:

         When the collection canister is full DO NOT operate the Vacuum Regulator.  

    

     The WARRANTY WILL BE VOIDED if the canister overfl ows and

    

     contaminates the Vacuum Regulator.

!

WARNING:

!

WARNING:

Summary of Contents for VR-CIU2-F2D

Page 1: ... 448 0770 Fax 847 855 6300 www ohiomedical com 25B East Pearce St Richmond Hill ON L4B 2M9 Canada Tel 905 764 7736 Fax 905 764 7743 www amvex com 0413 0413 Models Included DIGITAL D ANALOG A VR CIU2 F2D VR CIU2 F2A VR I2U2 F2D VR I2U2 F2A VR PIU2 F2D VR PIU2 F2A VR PPU2 F2D VR PPU2 F2A VR NIU2 F2D VR NIU2 F2A VR NNU2 F2D VR NNU2 F2A Rx ONLY ...

Page 2: ... Cover Screws 4 pcs 2 Mode Face Assembly 3 Mode Face Assembly 3 Mode 2 Mode Intermittent 17 Back Plate Screws 2 pcs 18 Bottom Plate Screw 19 Orifice 20 Regulating Module 23 Pediatric Relief Screw 24 Pediatric Relief Body 25 Pediatric Relief Ball 26 Pediatric Relief O ring 27 Intermittent Module 28 Chamber O ring 0 200mmHg Gauge Analog or Digital 0 300mmHg Gauge Analog or Digital 0 160mmHg Gauge An...

Page 3: ... to room temperature before use User Responsibility This device is to be used ONLY by people who have been properly trained on the operation of the device Operation of this device is not to be done if flammable anesthetics are present due to the possibility of explosion caused by static charge This product performs as explained in this manual as long as the assembly use repair and maintenance are ...

Page 4: ...tubing is recommended by NFPA National Fire Protection Association NFPA 99 2002 Healthcare facilities pages 497 498 To prevent possible contamination of the regulator a high flow suction filter or an overflow safety trap provided by Amvex is recommended between the regulator and the collection canister NFPA recommends use of an overflow trap to protect the vacuum regulator outlet and vacuum system...

Page 5: ...slowly to reach various vacuum rates Ensure that the level of vacuum maintains consistently when the vacuum tubing is kinked 5 Move the selector switch to INT 6 Kink vacuum tubing 7 Timing cycles are approximately 16 seconds on and 8 seconds off NOTE The intermittent unit starts in the off cycle 8 Decrease the vacuum level to zero and move the selector switch to the OFF position Pediatric Neonatal...

Page 6: ...body during timing NOTE The off time must be adjusted prior to the on time 6 The unit will begin in the off mode of the intermittent cycle To increase the off time turn the OFF Needle Valve clockwise To decrease the off time turn the OFF Needle valve counterclockwise 7 After the off time has been adjusted to the desired timing the on time may be adjusted To increase the on time turn the ON Needle ...

Page 7: ... mixture may cause small surface cracks to some of the plastic parts that may not be apparent to the user CAUTION Do not steam autoclave immerse in liquid or gas sterilize the Vacuum Regulators This may damage the unit CAUTION If Vacuum Regulator becomes contaminated internally warranty is voided Do not send Vacuum Regulator back to the manufacturer Follow your facilities procedures for handling c...

Page 8: ...operated under conditions of normal use regular periodic maintenance and service is performed and repairs are made in accordance with this operation manual The warranty period for all expendable parts of the Product is sixty 60 days from the date the Company ships the Product to the customer The foregoing warranty shall not apply if the Product has been repaired or altered by anyone other than the...

Reviews: