With the keyboard unlocked, and the pump displaying [OFF], the (P)
programming button will allow you to step through a menu to set the
following parameters in turn:

Press the (P) button to select the setting.

Use the (+) or (-) button to select required value.

Press the (P) button to confirm value and select next parameter.

OR

Press no button for 5 seconds and value will be set.

Start the pump by pressing the [+/ON] button.

The symbols [Pr] will appear on the screen to allow you to prime the
infusion set. Press and hold this button to prime up to 1mL. Repeat this
to give further primes if necessary.

Press 

 the [+/ON] button to start the infusion.

At the end of an infusion the plunger will automatically return to its
programmed start position, and the pump will switch off.

To stop the pump earlier press the [-] and [+/ON] keys together. The
word [Stop] will appear on the screen and will flash. If you wish to
terminate the infusion at this time press and hold the [-] button for 7
seconds or if you want to restart the infusion press the [+/ON] key.

If you stop the infusion before the pump has reached the end you will
need to reset the pump before starting the next infusion. With the
pump displaying [Stop] press the [P] and [-] keys together. The text
[END] will appear on the screen and the pump plunger will return to its
programmed start position.

Programming the pump

Setting procedure

Starting infusion

Ending infusion

SG005/19BM

t-mode

F-mode

End of infusion alarm

 On/Off

On/Off

Infusion Time/Rate

30 mins to 400 hrs

0.10 to 100 mL/hr

Infusion Volume

0 to 40 mL

0 to 40mL