Amsco RELIANCE 300 Installation Instructions Manual Download Page 5

1

Uncrating/Installation Instructions

122993-618

The following is a summary of the safety precautions which must be observed when uncrating, installing and
operating this equipment.  WARNINGS indicate the potential for danger to personnel and CAUTIONS indicate the
potential for damage to equipment.  These precautions are repeated (in whole or in part), where applicable,
throughout the instructions.

WARNING - PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD:

When moving the unit, use a forklift.  Do not use door(s) or door handle(s) for lifting.

WARNING - LACERATION/EYE INJURY HAZARD:

When removing bands, wear eye protection and always use a tool specifically designed to cut the bands.
The bands used to secure this carton can cause personal injury when cut and tension is released.

WARNING - LACERATION HAZARD:

When removing bolts, wear gloves to protect your hands.

WARNING - BURN HAZARD:

When checking the automatic stop feature, wear gloves and face protection and open chamber door slowly.
Hot water/steam may be sprayed through door opening if door is opened during a cycle.

When blowing out utility supply lines, wear gloves, face shield and protective clothing.

WARNING - PERSONAL INJURY HAZARD:

When closing chamber door, use door handle.  Keep fingers/hands away from top and sides of door to prevent
fingers from being crushed.

WARNING - FALL HAZARD:

To prevent falls, keep floors dry. Promptly clean up any spills or drippage.

CAUTION - POSSIBLE EQUIPMENT DAMAGE:

After utilities are connected to washer, slowly remove the protective adhesive paper from the exterior
cabinet panels to reduce the level of static discharge.

When removing adhesives from stainless steel, use a small amount of non-flammable cleaning solvent.
Rub in a back-and-forth motion (in same direction as surface grain).  Solvent rubbed in a circular motion
or applied with a wire brush or steel wool on door and chamber assemblies can be harmful to stainless steel.
Do not use solvents on painted surfaces.

Once three-phase power is connected, check pump for correct rotation.  Incorrect pump rotation may result
in pump damage and improper cleaning action.

Before operating unit, always position each manifold over a manifold connector.  If manifolds are not
positioned correctly, damage may result and unit will be unable to effectively wash loads.

1

SUMMARY OF WARNINGS AND CAUTIONS

Summary of Warnings and Cautions

Summary of Contents for RELIANCE 300

Page 1: ...UNCRATING INSTALLATION INSTRUCTIONS RELIANCE 300 Laboratory Glassware Washer 12 98 P 122993 618 Rev 1 ...

Page 2: ...AND HOLD CUSTOMER HARMLESS FROM ANY AND ALL CLAIMS LIABILITY DAMAGES OR EXPENSES DUE TO PERSONAL INJURIES INCLUDING DEATH TO EMPLOYEES OF STERIS AND CUSTOMER AND TO THIRD PARTIES AND FOR PROPERTY DAMAGE TO THE EXTENT OF THE WILLFUL ACTS OF THE NEGLI GENCE OF STERIS OR THE STRICT LIABILITY OF STERIS CAUSED BY THE ACTS OR OMISSIONS OF STERIS CUSTOMER AGREES TO DEFEND INDEMNIFY AND HOLD STERIS HARMLE...

Page 3: ...stalling and starting up this equipment contact the STERIS Regional Office nearest you The Reliance 300 Laboratory Glassware Washer is designed for efficient cleaning of laboratory glassware plasticware and metal goods typically used in research support production and quality control laboratories This washer is specifically designed to only processgoodsasoutlinedintheOperatorManual P 122993 619 If...

Page 4: ...LE TITLE TITLE TITLE TITLE PAGE PAGE PAGE PAGE PAGE 1 SUMMARY OF WARNINGS AND CAUTIONS 1 2 UNCRATING INSTRUCTIONS 3 Open Carton 3 Move Washer and Remove Skid 4 Prepare Washer for Installation 4 3 INSTALLATION INSTRUCTIONS 5 Before Installing Equipment 5 Connect Utilities 5 Cleanup 6 4 INSTALLATION CHECKLIST 7 5 OPERATIONAL TEST 9 ...

Page 5: ...g out utility supply lines wear gloves face shield and protective clothing WARNING PERSONALINJURYHAZARD Whenclosingchamberdoor usedoorhandle Keepfingers handsawayfromtopandsidesofdoortoprevent fingers from being crushed WARNING FALLHAZARD To prevent falls keep floors dry Promptly clean up any spills or drippage CAUTION POSSIBLEEQUIPMENTDAMAGE After utilities are connected to washer slowly remove t...

Page 6: ...ve Packing Material Figure 2 Remove Packing Material Figure 2 Remove Packing Material P o l y s t y r e n e Wrap Cardboard Corners Bands S k i d Uncrating Instructions Figure 3 Reliance 300 Laboratory Glassware Washer Figure 3 Reliance 300 Laboratory Glassware Washer Figure 3 Reliance 300 Laboratory Glassware Washer Figure 3 Reliance 300 Laboratory Glassware Washer Figure 3 Reliance 300 Laboratory...

Page 7: ... until after utilities are connected 10 Locate keys for the lower service access panel 11 Remove lower service access panel and check tip indicator located inside lower service compartment see Figure 3 Tip indicator contains a blue compound at the bottom of the indicator If unit has been tipped residue from the blue compound will be found higher up in the indicator If unit was tipped notify STERIS...

Page 8: ...tective adhesive paper from the exterior cabinet panels until after utilities are connected 1 Open chamber door If accessories are ordered with unit remove nylon straps securing accessories to chamber frame and slide accessories out of wash chamber 2 Remove Operator Manual from wash chamber Make sure appropriate person receives Operator Manual 3 Remove nylon straps securing rotary spray arm assemb...

Page 9: ...phase 3 wire power Check Equipment Drawing or Identification Plate located inside lower service compartment on bottom of chamber sump see Figure 3 for proper voltage Blow out all utility supply lines to remove chips scale etc 2 Make sure washer is placed as shown on Equipment Drawing in correct relation to building supply lines If unit is not at installation site refer to Uncrating Instructions Se...

Page 10: ...lower service access panel 2 Remove all protective paper from the washer cabinet panels Slowly peel paper away from stainless steel to reduce the level of static discharge 3 Remove any adhesive found on panels door s and inside chamber with a small amount of non flammable cleaning sol vent Keep solvent away from all painted surfaces or damage may result 4 Inspectwasherandworkareatobesureallmateria...

Page 11: ... shown on Equipment Drawing with re quired clearance space and in relation to building supply lines Washer is level If washer is equipped with Drain Discharge Cooldown and Cold Water Pre wash option building cold water line supplies water to unit at dynamic pressure and temperature as specified on Equip ment Drawing Building hot water line supplies water to unit at dynamic pressure and temperature...

Page 12: ...er door Make sure main power switch located in upper service compartment see Figure 3 is in ON position 6 Press Power Off Standby Touch Pad once see Figure 6 Verify that model name temporarily appears on the display screen RELIANCE MODEL 300 followed by the first cycle menu screen LIGHT MEDIUM HEAVY CYCLE 4 Indicates flashing position 7 Press Cycle Start touch pad once to select the LIGHT cycle Ve...

Page 13: ...tank does not overfill 10 Once chamber sump is filled with water check rotation of the recirculation pump as follows a Unlock and remove lower service access panel b While viewing the pump motor shaft alternately press the Cycle StartandStop Resettouchpads c Verify motor shaft is rotating in direction indicated by the arrow onthemotorhousing Ifpumpisrotatinginthewrongdirection disconnect main 3 ph...

Page 14: ... and allow chamber and accessories to cool before unloading Verify display screen returns to the first cycle menu once door is opened 22 Remove screens from utility supply line strainers and check for debris Clean screens if necessary WASHERISNOWREADYFORNORMALOPERATION Contact your local STERIS representative for a demonstration on how to operate the equipment Refer to the Operator Manual P 122993...

Page 15: ...12 122993 618 Uncrating InstallationInstructions NOTES ...

Page 16: ...e precise maintenance program and factory trained technicians to assure you of maximum productivity Our STERIS service technicians thoroughly inspect clean adjust and provide all necessary maintenance to keep your equipment performing according to factory speci fications all at an established economical rate that you can plan for We have more than 5 000 customers who are benefiting from STERIS mai...

Reviews: