7-3
Cycle and Control Value Programming
Operator Manual
P129394-176
Cycle values (sterilize temperature and time, and dry time) are
adjustable within the limits shown in
.
•
TEMP:
For changing sterilize exposure temperature within the
ranges shown.
•
STER:
For changing the sterilization exposure time. Use
number touchpads to enter a new value. Left and right arrow
touchpads can be used to cursor through an existing value.
•
DRY:
For changing the dry time.
NOTE:
To return cycle values to original state, press RESTORE ALL
DEFAULT VALUES. This touchpad can also be used to cancel
previous changes to a cycle.
TEMP
STER
DRY
NAME
SELECT NEW DRY TIME
OR SELECT ANOTHER ITEM TO CHANGE
RANGE = 0-99 MINUTES
13
1
2
3
RESTORE ALL
DEFAULT VAL-
UES
4
5
6
7
8
9
EXIT
<-
0
->
Table 7-1. Adjustment Ranges for Cycle Values
Cycle Type
Sterilize Time
Sterilize Temperature
*
* Sterilize time limits:
»
Prevac or SFPP cycles set to a temperature of 270-274°F (132-134°C) are limited to a four-minute
minimum sterilize time.
»
Prevac or SFPP cycles set to a temperature of 275°F (135°C) are limited to a three-minute minimum
sterilize time.
Dry Time
270°F (132°C) Prevac
4 - 99 minutes
270 - 274°F (132 - 134°C)
1 - 99 minutes
275°F (135°C) Prevac
3 - 99 minutes
275°F (135°C)
270°F (132°C) SFPP
4 - 99 minutes
270 - 274°F (132 - 134°C)
0 - 99 minutes
275°F (135°C) SFPP
3 - 99 minutes
275°F (135°C)
250°F (121°C) Gravity
30 - 99 minutes
250 - 269°F (121 - 131°C)
270°F (132°C) Gravity
3 - 99 minutes
270°F (132°C)
250°F (121°C) Liquid
45 - 99 minutes
250°F (121°C)
N/A
Test Cycles
Dart (Bowie-Dick Test)
Dart Warm-Up
Leak Test
Values Cannot Be Adjusted