4-7
Techniques of Sterilization
Operator Manual
P129394-176
4.7 Techniques of
Sterilization for
Liquid Cycle
Important:
The liquid cycle is for non-patient contact use only.
for recommended Liquid cycle parameters. The
recommended times indicated in
assume the use of
vented bottles. Minimum recommended sterilization time (fourth
column) includes the time required to bring the solution up to the
sterilize temperature plus the time required to achieve sterilization.
Table 4-3. Liquid Cycle Processing Guidelines
WARNING – EXPLOSION
HAZARD: This sterilizer is
not designed to process
flammable compounds.
WARNING – PERSONAL
INJURY HAZARD: Avoid
personal injury from
bursting bottles. Liquid
sterilization cycle must
only be used for liquids in
borosilicate (Pyrex) flasks
with vented closures.
WARNING: The liquid
cycle is for non-patient
contact use only.
WARNING – BURN
HAZARD: When sterilizing
liquids, you must observe
the following procedures:
•
Use Liquid cycle only.
•
Use only vented closures.
•
Use only Type I
borosilicate glass bottles.
•
Do not allow hot bottles to
be jolted.
• Steam may be released
from the chamber when
door is opened. Step
back from the sterilizer
each time the door is
opened to minimize
contact with steam
vapor.
CAUTION – POSSIBLE
EQUIPMENT DAMAGE:
Sterilization of chloride-
containing solutions (e.g.,
saline) can cause chamber
corrosion and is not
recommended by the
manufacturer. If, however,
chloride-containing
solutions must be
processed, clean the
chamber after each use.
!
!
!
!
!
Sterilizer Size
Number of
Containers
for Full
Load
Volume of
Liquid in
One
Container
Minimum
Recommended
Sterilize Time
at 250°F
(121°C) in
minutes
16 x 16 x 26"
(406 x 406 x 660 mm)
15
1000 mL
45
20 x 20 x 38”
(508 x 508 x 965 mm)
with Loading Car
32
20 x 20 x 38”
(508 x 508 x 965 mm)
with Rack and Shelves
28