Operator Manual
Cycle and Control Value Programming
7.5 Change Machine
Setup
All changes are made to displayed settings using touch screens. No
mechanical adjustments to the sterilizer are necessary.
Generally, the Setup options are used to change the way the
sterilizer operates. The control has an Access Code Security feature.
If the Access Code is enabled, all or some of these options can be
secured or “locked out” by the supervisor (see Access Code later in
this section for more information on Access Code). All Setup options
are accessed from the Setup Menu.
A summary of the setup options that can be adjusted are listed in
. Each value is detailed in this section.
Table 7-2. Change Machine Setup
MACHINE SETUP
Once the access code has been set, all cycles values and Change Values options can be
selected to lockout, making them unavailable for change by an operator without entering the
Access Code.
This setting is used to secure any of the Change Values options under the access code.
Once a function is locked out, the Access Code must be entered before the setting can be
accessed or changed.
This setting permits the operator to program the sterilizer to automatically shut off its steam
and water at the end of the work day, to conserve utilities. Shut down and power-up times
can be programmed for any time of the day on Weekdays or the Weekend.
This is used to assign a six-character, alphanumeric code to the sterilizer. This code appears
in the heading of all printouts.
This setting allows the control to display times using either AM/PM format or 24 hour format.
This setting allows the sterilizer to print a condensed version of the cycle printout to conserve
on paper usage.
This option allows adjustment of ALARM, END OF CYCLE and TOUCHPAD signals tones.
Tones can be independently adjusted to one of three volume levels. ALARM signal tone can-
not be turned off.
This is used to select the temperature and pressure units displayed and printed out by the
sterilizer. The sterilizer is capable of displaying temperature as either Fahrenheit or Celsius;
pressure can be displayed as psig/inHg, millibar or psia.
The sterilizer can be programmed to change the way the date is displayed and printed. The
normal setting is to display Month/Day/Year (M/D/Y); but this can be changed to
Year/Month/Day (Y/M/D), or Day/Month/Year (D/M/Y).
Sterilizer can be set to automatically furnish a duplicate printout of each cycle at the end of
the cycle. First line will always read — DUPLICATE PRINT — and complete printout of cycle
data will be furnished.
Summary of Contents for 36 H
Page 2: ......
Page 14: ...xii P129394 068 Operator Manual Techniques of Sterilization LIST OF FIGURES Descrption Page ...
Page 59: ...6 11 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Page 63: ...6 15 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Page 67: ...6 19 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Page 78: ...6 30 P129394 068 Operator Manual Sterilizer Operation Healthcare Capital Equipment ...
Page 116: ...8 10 P129394 068 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Page 140: ...9 24 P129394 068 Operator Manual Troubleshooting Healthcare Capital Equipment ...
Page 152: ...10 12 P129394 068 Operator Manual Service Procedures Healthcare Capital Equipment ...