![Amsco 36 H Operator'S Manual Download Page 84](http://html1.mh-extra.com/html/amsco/36-h/36-h_operators-manual_2936188084.webp)
7-6
P129394-068
Operator Manual
Cycle and Control Value Programming
F
IGURE
, below, is a continuation of the flow chart in
F
IGURE
.
Figure 7-2. Change Cycle Values, Part 2 of 3
F
IGURE
, below, is a continuation of the flow chart in
F
IGURE
.
Comments:
Important:
Pressing
EXIT
causes the sterilizer to leave
Change Values, saving any
changes made to the point
where
EXIT
was pressed.
7-3
If cycle is locked out, Access
Code Screen (#37) appears.
The access code must be
entered using the following
procedure:
•
If the code is known, enter
it using the touch screen
key pad. Once the entire
four digit code has been
entered, press the
ENTER
touch screen pad. The
access code digits are not
displayed.
If the code is not known, press
MENU
to return to the
previous screen.
A printout of cycle values and
the cycle count is provided by
pressing
PRINT VALUES
.
Any changed cycle values
and the cycle count are
automatically printed after
returning to screen #10 from
screen #11.
Summary of Contents for 36 H
Page 2: ......
Page 14: ...xii P129394 068 Operator Manual Techniques of Sterilization LIST OF FIGURES Descrption Page ...
Page 59: ...6 11 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Page 63: ...6 15 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Page 67: ...6 19 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Page 78: ...6 30 P129394 068 Operator Manual Sterilizer Operation Healthcare Capital Equipment ...
Page 116: ...8 10 P129394 068 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Page 140: ...9 24 P129394 068 Operator Manual Troubleshooting Healthcare Capital Equipment ...
Page 152: ...10 12 P129394 068 Operator Manual Service Procedures Healthcare Capital Equipment ...