![Amsco 36 H Operator'S Manual Download Page 37](http://html1.mh-extra.com/html/amsco/36-h/36-h_operators-manual_2936188037.webp)
5-5
Component Identification
Operator Manual
P129394-068
5.4 Alarm Displays
Alarm displays tell operators and technicians when the sterilizer is
experiencing an abnormal condition. Alarm conditions can be
caused by failure of utility supplies or sterilizer components.
, details the steps an operator can take to solve
most alarm conditions. Typical alarm displays are shown in
Figure 5-5
.
When an alarm occurs during cycle operation, a display appears on
the screen, accompanied by an audible tone. This display indicates
the problem as determined by control sensors, and lists a brief
troubleshooting list. The operator should follow the instructions on
the screen, if possible. If these instructions fail to clear the alarm,
consult your departmental supervisor or a trained service technician
before using the sterilizer further.
Figure 5-5. Typical Alarm Displays
STATUS . . ALARM
205
TOO LONG IN CHARGE
CHAMBER . . 158.0F 0psig
STERILIZER WILL:
AUTOMATICALLY TRY TO COMPLETE CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CHECK STEAM SUPPLY VALVE
->IF CLOSED, OPEN VALVE
3. IF ALARM RECURS, CALL SERVICE
SILENCE
STATUS
PAPER
SERVICE
ALARM
FEED
HELP
STATUS . . ALARM
206
TOO LONG IN CHARGE
CHAMBER DID NOT REACH STERILIZE
TEMPERATURE WITHIN ALLOTTED TIME
CAUSES AND CORRECTION:
1. STEAM PRESSURE LESS THAN 50 PSIG
->CHECK STEAM SUPPLY PIPING
2. STEAM REGULATOR MALFUNCTION
->REPAIR
2. SOLENOID VALVE MALFUNCTION
->REPAIR S09
->REPAIR S02
3. CONTROL OUT OF CALIBRATION
->RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
PAPER
FEED
EXIT
ABORT
Alarm Notification Screen
Alarm Service Information Screen
Summary of Contents for 36 H
Page 2: ......
Page 14: ...xii P129394 068 Operator Manual Techniques of Sterilization LIST OF FIGURES Descrption Page ...
Page 59: ...6 11 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Page 63: ...6 15 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Page 67: ...6 19 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Page 78: ...6 30 P129394 068 Operator Manual Sterilizer Operation Healthcare Capital Equipment ...
Page 116: ...8 10 P129394 068 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Page 140: ...9 24 P129394 068 Operator Manual Troubleshooting Healthcare Capital Equipment ...
Page 152: ...10 12 P129394 068 Operator Manual Service Procedures Healthcare Capital Equipment ...