1-2
150830-026
Operator Manual
Safety Precautions
WARNING – PERSONAL INJURY HAZARD:
Healthcare professionals must ensure patients are positioned and monitored to prevent compromising respiration,
nerve pathways, or circulation.
When installing any table accessory, check for correct attachment and tighten securely (if appropriate). Do not
use worn or damaged accessory. Check installation before using any accessory.
There is a 1,000-lb (452-kg) patient weight limit if patient is in normal orientation and a 500-lb (226-kg) patient weight
limit if patient is in reversed orientation; however, the accessory load rating may be lower. Do not exceed the
accessory load rating if it is lower than the table rating.
Unanticipated table movement could cause patient injury. Patient must be secured to the table in accordance with
recommended positioning practices.
Do not immerse any part of foot control in liquids; this could cause unanticipated table movement, leading to patient
injury. Always cover control with a plastic bag before using.
If the integrity of the external protective earth conductor installation or arrangement is in doubt, operate the table
from its internal power source.
WARNING – INSTABILITY HAZARD:
Stabilize table when transferring patient.
Possible patient or user injury, as well as table or accessory failure, may result from using STERIS table accessories
for other than their stated purpose - or from using, on STERIS tables, accessories manufactured and sold by other
companies.
Patient Transfer Board must be used as a leg support only. It is not intended to support upper body weight of a
patient.
WARNING – PINCHING AND TIPPING HAZARD:
Patient injury may result if the operator of this table is not completely familiar with the controls for patient positioning
and table operation.
WARNING – PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD:
Safe and reliable operation of this equipment requires regularly scheduled preventive maintenance, in addition to
the faithful performance of routine maintenance. Contact STERIS to schedule preventive maintenance.
Repairs and adjustments to this equipment must be made only by fully qualified service personnel. Nonroutine
maintenance performed by inexperienced, unqualified personnel or installation of unauthorized parts could cause
personal injury, invalidate the warranty, or result in costly damage. Contact STERIS regarding service options.
WARNING – INFECTION HAZARD:
To protect against aerosols being reflected from contaminated surfaces, wear rubber or plastic gloves, masks and
eye protection and follow OSHA blood-borne pathogens standards when cleaning.
WARNING – DISPOSAL HAZARD:
This product contains materials which may require disposal through appropriately licensed and permitted
hazardous waste management firms.