TMF882X
Using the EVM Software
Demo Kit Manual
• PUBLIC
UG000495
• v4-00 • 2021-Dec-01
33
│ 26
Calibration Handling
Control the EVM GUI Calibration Handling.
Figure 25:
Command Overview
Command Code
Parameter
Example
C
0 - Initiate new calibration,
1 - Clear calibration,
2 - Get calibration data
The EVM GUI ignores all other
parameter values
(C2)
3.6.3
Automation Example Session
To operate the following example, the EVM GUI needs to be started first. Start the Windows shell
"cmd" and connect through the Windows program
“telnet”:
●
telnet 127.0.0.1 39998
This causes an immediate response and in the GUI the status ‘Automation On’ is displayed (removed
headers in the output data):
sep=;
#VER;ToF Serial Number;App0 Version Number;GUI Version Number;Host Driver Version
Number;Chip ID;Chip Revision
#VER;4c-c0-00-00;2.49.4.2;2.4.23.0;2.6.2.19;8;0
#CAL;1594816397308;0;2;00;0B;03;00;01;FF;26;02;79;56;34;12;53;49;47;F7;6B;A2;07;00
;6B;A2;07;00;6B;A2;07;00;6B;A2;07;00;6B;A2;07;00;6B;A2;07;00;6B;A2;07;00;6B;A2;07;
00;6B;A2;07;00;6B;A2;07;00;77;09;02;00;02;00;00;00;02;00;00;00;02;00;00;00;02;00;0
0;00;02;00;00;00;02;00;00;00;02;00;00;00;02;00;00;00;02;00;00;00;00;00;00;00;00;00
;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;
00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;0
0;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00;00
;00;00;00;00
Send the command
(m0)
- it executes one measurement and transfers the enabled histograms
(removed all but one histogram).
●
(m0)
#TMP;1594816405200;207;33
#OBJ;1594816405203;207;3;3;1970;19;0;0;1938;42;0;0;1961;16;0;0;1947;49;0;0;1931;11
8;0;0;1944;47;0;0;1946;20;0;0;1929;38;0;0;1948;20;0;0;0;0;0;0;0;0;0;0;0;0;0;0;0;0;
0;0;0;0;0;0;0;0;0;0;0;0;0;0;0;0;0;0;0;0;0;0
#LIGHT;1594816405205;207;25079;199532;14146