- 27 -
c
omplete
f
Inal
I
nflate
/d
eflate
t
est
1 . After all components are connected, completely inflate and
deflate the cylinders at least once to ensure the device is
functioning properly; to check the quality of the erection;
and to evaluate flaccidity .
Note: The erect penis should present a satisfactory cosmetic result.
Note: The flaccid penis should lie close to body when deflated.
There may be some swelling that precludes a good flaccid result.
Note: If erect or flaccid results are not acceptable, check the amount
of fluid in reservoir and adjust the volume if necessary.
2 . Before ending the procedure, squeeze the deflation button to
allow the cylinders to partially deflate so that some fluid still
remains in the cylinders postoperatively . This will ensure the
cylinder capsules are large enough to prevent resistance to
inflation .
To prevent autoinflation, squeeze the deflation button as the last
action before closing the incision .
3 . Close the incision .
Note: Some physicians close the dartos in two layers with running
2-0 chromic catgut suture then close the skin.
4 . Apply a wound dressing and leave partially inflated .
5 . Tape penis to abdomen
(Figure 4-24).
6 . Optionally, a drain may be placed for 12-24 hours
Figure 4-24. Tape Penis to Abdomen
Surgical Procedures
(continued)