Amprobe FG2C-UA User Manual Download Page 12

6

 

(Sweep On/Off): Turn clockwise the knob clockwise for MAX frequency and counterclockwise for MIN 
frequency. (Keep the knob pointer within the scale range on the panel). Pull out the knob to start the auto 
sweep operation; the upper frequency limit is determined by the knob position.

14   SWEEP TIME (FG3C-UA)

 

Control and LIN/LOG Selector:

 

(1)   Rotate the knob clockwise to adjust sweep time for MAX, or counterclockwise for MIN.

 

(2)   Select Linear sweep mode by pushing in the knob or LOG sweep mode by pulling out the knob.

15.   MOD ON/OFF Selector (FG3C-UA)

 

Press the button once, the indicator will light, and the output will be modulated by internal 400Hz sinewave 
or press the button again, the indicator will be off, and the output will be modulated an external signal via 

VCF/MOD 

in connector 

(21) 

.

16.   MOD/DEPTH (FG3C-UA)

 

SWEEP RATE and AM/FM Selector:

 

(1) Sweep width can be controlled from 0 to 100 ratio.

 

(2) Adjust modulation ratio by turning the knob clockwise for MAX, or counterclockwise for MIN.

 

(3) Select AM(amplitude modulation) mode by pushing in the knob or FM(frequency modulation) mode 
by pulling out the knob.

17   MOD EXT (FG3C-UA)

 

Selector: Press the button once, the indicator will light, and EXTernal MODulation is selected. Press the button 
again, the indicator will be off, and internal MODulation is selected.

18.   INT/EXT Counter Switch (FG3C-UA)

 

Select internal counter mode (count the frequency of the FG3C-UA output) or select EXT counter mode for an 
independent counter (input signal from 

BNC (19) termina

l).

19.   EXT. Counter Input Terminal (FG3C-UA)

 

Accepts external signals for measurement.

20.   TTL/CMOS Output Terminal

 

TTL/CMOS compatible signal output

21.   VCF/MOD

 

Input Terminal: Connector for the input voltage required to perform the “voltage control frequency” 
operation or the EXT modulation operation.

22.   Main Output Terminal

 

Main signal output.: The Output signal is calibrated for a 50 

 load.

23.   GCV Output (FG3C-UA)

 

This is a DC voltage output that flows the output frequency of each range. Minimum (~ 0 VDC) to maximum 
(~ +2 VDC) 

24.   115/230 Switch 

 

This switch selects the mains power voltage.

USAGE PROCEDURES 

The following section describes the basic setup and controls for operating the functions of this instrument.
One of the best ways to observe waveforms is to connect the function generator to an oscilloscope. Observe the 
effects that different controls have on the waveforms displayed on the oscilloscope when preceding with the 
following steps:

First-step check:

(1)  Ensure that the 

mains voltage switch (24) 

on the rear panel of this instrument is compatible with the mains 

supply.
Connect the instrument to the mains supply using the power cord supplied.
(2)  Press 

PWR (1) 

switch, ensure all the rotary knobs are pushed in, and rotate the 

AMPL (12) 

knob to fully 

counterclockwise.
(3)  Rotate the 

FREQ (13) 

control fully counterclockwise.

Summary of Contents for FG2C-UA

Page 1: ...User Manual Mode d emploi Bedienungshandbuch Manuale d Uso Manual de uso FG2C UA FG3C UA Function Generator...

Page 2: ......

Page 3: ...FG2C UA FG3C UA Function Generator Users Manual Mode d emploi Bedienungshandbuch Manuale d Uso Manual de uso PN 2099447 April 2007 Rev 2 2007 Amprobe Test Tools All rights reserved Printed in China...

Page 4: ......

Page 5: ...tar libre de defectos de material y mano de obra durante 1 a o a partir de la fecha de adquisici n Esta garant a no cubre fusibles bater as descartables o da os que sean consecuencia de accidentes neg...

Page 6: ......

Page 7: ...from ground and never touch exposed wiring connections test probe tips or any live circuit conductors Do not operate instrument in an explosive atmosphere flammable gases fumes vapor dust Do not use t...

Page 8: ...er and disposed of immediately as a plug with bared wires is hazardous if engaged in a live socket Any rewiring must be carried out in accordance with the information detailed on this label PREPARATIO...

Page 9: ...P O W E R CM O S M IN Fig 1 FG2C UA Front Panel Fig 2 Rear Panel...

Page 10: ...M k m H z P O W E R S W E E P WIDTH L I N A M F M D U T Y C M O S M I N M O D 1 M M O D Fig 3 FG3C UA Front Panel Fig 4 FG3C UA Rear Panel...

Page 11: ...sh button on the panel as shown in Table 1 Table 1 Push Button Frequency Range 1 0 3Hz to 3Hz 10 3Hz to 30Hz 100 30Hz to 300Hz 1k 300Hz to 3kHz 10k 3kHz to 30kHz 100k 30kHz to 300kHz 1M 300kHz to 3MHz...

Page 12: ...EXT Counter Switch FG3C UA Select internal counter mode count the frequency of the FG3C UA output or select EXT counter mode for an independent counter input signal from BNC 19 terminal 19 EXT Counte...

Page 13: ...ramp waveform 5 Pull out the AMPL 12 knob to attenuate the output signal by 20dB TTL CMOS signal output 1 Select the Range 7 and rotate the FREQUENCY 13 knob to set the frequency 2 Connect BNC connec...

Page 14: ...om 0 to 1 Vpp The FG3C UA output signal amplitude is set by the FG3C UA amplitude controls Note The sweep width can only be adjusted during the sweep cycle and it cannot be stopped AM FM Modulation in...

Page 15: ...e rear of the Function Generator Warning Failure to turn off and remove power from the Function Generator before changing the mains power fuse could result in damage to the instrument and the operator...

Page 16: ...check the Where to Buy section on www amprobe com for a list of distributors near you European Correspondence Address Amprobe Test Tools Europe P O Box 1186 5602 BD Eindhoven The Netherlands Correspon...

Page 17: ...N A 0 to 5 Modulation Frequency N A 400Hz INT DC to 20 kHz EXT EXT Sensitivity N A 10Vpp for 10 modulation 12 Frequency Counter Int Ext N A Switch selector Range N A 0 3 Hz to 3 MHz 5 Hz to 150 MHz E...

Page 18: ...CE Marking However electrical noise or intense electromagnetic fields in the vicinity of the equipment may disturb the measurement circuit Measuring instruments will also respond to unwanted signals...

Reviews: