11
6) SPECIFICATIONS
GENERAL SPECIFICATIONS
Display :
3-3/4 digits 4000 counts LCD display(s)
Update Rate :
3 per second nominal
Polarity :
Automatic
Low Battery :
Below approx. 2.4V
Operating Temperature :
32
!
F TO 104
!
F (0
!
C TO 40
!
C)
Relative Humidity :
Maximum relative humidity 80% for temperature up to 31
!
C decreasing
linearly to 50% relative humidity at 40
!
C
Altitude :
Operate below 6,562ft (2000m)
Storage Temperature :
-4
O
F to 140
º
F (-20
O
C to 60
O
C), < 80% R.H. (with battery removed)
Temperature Coefficient :
nominal 0.15 x (specified accuracy)/
O
C @ 32
!
F - 64
!
F OR 82
!
F
- 104
!
F (0
O
C -18
O
C or 28
O
C -40
O
C), or otherwise specified
Sensing :
Average sensing for ACD-6 PRO; True RMS for ACD-6 TRMS PRO
Safety :
Meets IEC61010-2-032(1994), EN61010-2-032(1995), UL3111-2-032(1999).
Category III 600 Volts ac & dc.
Transient protection :
6.5kV (1.2/50
"
s surge) for all models
Pollution degree :
2
E.M.C. :
Meets EN61326(1997, 1998/A1), EN61000-4-2(1995), and EN61000-4-3(1996)
In an RF field of 3V/m:
Total Accuracy = Specified Ac 45 digits
Performance above 3V/m is not specified
Overload Protections :
ACA Clamp-on jaws : AC 1000A rms continuous
+
& COM terminals (all functions) : 600VDC/VAC rms
Power Supply :
One 3V coin battery (ANSI/NEDA-5004LC, IEC-CR2032)
Power Consumption :
ACD-6 Pro: 2.2 mA typical
ACD-6 TRMS Pro: 2.8mA typical
APO Timing :
Idle for 30 minutes
APO Consumption :
40
"
A typical on all model functions except that 190
"
A typical on ACD-
6 TRMS PRO voltage & current functions
Dimension :
L224mm X W78mm X H40mm
Weight :
220 gm approx
Jaw opening & Conductor diameter :
45mm max
Accessories :
Test leads (pair), batteries installed, user's manual, & soft carrying pouch