9
button can quickly scroll the logged items. The meter gives short beeps when the last (or
he first) item is reached.
t
Note:
1.
When the meter’s memory is full or the meter battery is low, the meter will stop (end) the
Hi-Lo logging session and go back to the normal measuring mode.
2.
The data is stored to the non-volatile memory shortly after every one-minute item interval
to maximize data safety. However, the end-of-data sign can be stored only after the Hi-Lo
logging session is ended with a session stop “
”. Therefore, always stop “
” the Hi-Lo
logging properly before sliding the slide-switch function-selector to the
next
function position.
3.
After the Hi-Lo logging session hais ended, you can switch off the meter for
transportation, storage, or even battery changing. The logged data can also be downloaded
to PC computers through the optional purchase
RS232 KIT2
interface kit. Also see RS232C
C computer interface capabilities section.
P
Auto feature
The
Auto
(Automatic-Hold) feature automatically captures and displays significant stable
readings. Press and hold the
Auto
button for one second or more to toggle to the
Auto
ode. The LCD annunciators “Auto” & “ ”
turn on.
m
WARNING
To avoid electric shock hazard, do not use the
Auto (
Automatic-Hold) mode to determine
if a circuit is live. Unstable readings will not be captured and displayed.
HOLD feature
The Hold feature freezes the display for later viewing. Press the
HOLD
button
momentarily to toggle to the Hold mode. The annunciator “ ” turns on.
www.
.com