Ample Scientific Champion S-33 Manual Download Page 1

1. Introduction  

Thank  you  for  choosing  Champion  S-33  centrifuge  from  Ample  Scientific. 
Champion S-33 features: 

 

Durable and sophisticated design 

 

Temperature controlled motor to extend centrifuge life 
expectancy  

 

Temperature controlled inner chamber through vortex 
ventilation system providing stable centrifuging environment 

 

Stainless steel inner chamber 

 

Ultra smooth acceleration curve 

2. Product Structure 

 

 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

       

3. Installation and Setup 

1.

 

Inspect the centrifuge for any damages. Contact distributor for 
questions regarding damages. 

2.

 

Place centrifuge on steady and leveled surface. The area should be 
well ventilated, and the centrifuge should be kept at least 300mm 
away from surrounding objects in all directions. S-33 is for indoor 
use only. 

3.

 

Open the centrifuge lid: 

a.

 

Connect centrifuge to power source and turn on power 
switch from the back of the centrifuge. When panel lights 
up, press [Off] for the lid to automatically unlock. Listen for 
the ‘click’ sound before opening the lid. If the centrifuge 
does not unlock, proceed to [b.]. 

b.

 

Using the provided screw driver, insert into small opening 
on the right side of the centrifuge just below the lid; 
maneuver until a ‘click’ is heard and the lid is unlocked. 

4.

 

Remove packaging material in the inner chamber 

5.

 

Place buckets into rotor. Place circular rubber inserts into rotor 
buckets. Remove any debris or pieces of left over packaging 
material from inner chamber. 

6.

 

Close lid and make sure it is locked. Select time for 1 minute. Press 
[On] to start operation. 

7.

 

Make sure the whirring sound produced is smooth. If there is 
excessive vibration or abnormal sounds, stop the centrifuge 
immediately. Make sure rotor is correctly installed and there is no 
debris in centrifuge. If vibration and sound persist, please contact 
your distributor immediately. 

8.

 

Your centrifuge is now ready to be used. 

 

 

 

Figure 1 - Product External Structure 

Rotor 

Lid 

Centrifuge Chamber 

Speed Control 
and Display 

[On] button to start 
centrifugation 

Timer control 
and display 

Pg | 2 

Manual unlock opening 

[Off] button to stop 
centrifugation and to 
open lid 

Power switch is 
located in the 
back 

Pg | 1 