ASS190 Slide Stainer
AMOSASS190SM20150814-01
---1---
ASS190 Slide Stainer
Service Manual
1
:
The Service Manual of the instrument is compiled according to
its structure and characteristic.
2
:
The instrument provides the design of AC110V/220V, composes of tank, temperature controlling and engine movement.
1
、
Tank
1
)
:
2 pieces of water tank.
2
)
:
17 pieces of reagent tank.
2
、
Engine movement
1
)
:
The motor controls the up and down movement step-by-steps
.
2
)
:
The motor controls the left and right movement step-by-steps
3
:
1
2
3
4
5 6
7
8
9 10
11 12
13
14 15 16 17
18
19
Water Water
Hcl-
alcohol
Hematoxylin Eosin
Dimethyl
benzene
alcohol
Dimethyl benzene
100%
95%
85%
70%
75%
85% 95%
100
%
Note:
The Beginning tank is from Tank 6 & Tank 7 (the basket is staying in tank 6 to waiting to run
the program 1 and other basket in tank 7 wait to run program 2). Press key
RUN
, the system will test the
basket in the beginning tank automatically or not, if the two baskets all stay well in the Tank 6 and 7,
the instrument will run the program 2 firstly, and then run the program 1. The system provides eight
programs be edited. (Note: when programming, make sure the basket in tank 6 running don’t overtake
the basket in tank 7 to avoid getting stuck, if meet the stuck, the long buzzer rings will remind user. So,
PLEASE NOTE:1. Staying time of basket in the tank 6 is more than tank 7, the others time
General Provisions
Configuration
Distribution of Liquid Tank