User Manual for AML Oceanographic’s Smart SV&T
5
Using the Instrument
Pre-Deployment Procedures
•
4 to 6 weeks ahead
o
Use the receiving checklist to verify the instrument is in good working order
o
Verify the calibration is valid for the duration of the deployment. If it is not, ship the
instrument or Xchange sensors back to an authorized service centre for re-
calibration, or request an Xchange sensor replacement.
o
Lubricate the connector with silicone spray
•
Before leaving the jetty
o
Connect the instrument to a computer and power supply using the data cable
o
Place the instrument in fresh water ensuring no bubbles are trapped on the SV
transducer or reflector plates. The instrument should return sound speed readings
in agreement with the chart on page 3.
Configuring Sampling Parameters using SeaCast
AML Oceanographic’s SeaCast application software greatly simplifies the process of setting up an
instrument to collect data. SeaCast does not require instrument configuration files and is able to
generate copies of sensor calibration certificates on demand.
Full details on the instrument configuration process can be found in the SeaCast manual. Below
please find a quick summary of that process:
Selecting an Instrument for Configuration
On the Instrument Tab, the first row of fields ‘Port’, ‘Baud Rate’ and ‘Status’ control and display the
communications with the instrument.
Summary of Contents for Smart SV&T
Page 1: ...Smart SV T User Manual Document Version 2 1 Dated 03 February 2010...
Page 24: ...User Manual for AML Oceanographic s Smart SV T 23 Wiring Diagram RS 232 Communications...
Page 25: ...User Manual for AML Oceanographic s Smart SV T 24 RS 485 Communications...
Page 26: ...User Manual for AML Oceanographic s Smart SV T 25 Mechanical Drawings...
Page 27: ...User Manual for AML Oceanographic s Smart SV T 26...