Shipping & Receiving
Receiving an Instrument
When receiving an instrument, perform the following steps to ensure the instrument will be ready for
deployment when required:
•
Inspect the shipping container, looking for signs of damage. Damage to the shipping
container could indicate damage to the instrument inside.
•
The shipping package contains the following items, identified above:
o
Base•X instrument
o
Data•Xchange (if purchased)
o
Data/Power cable
o
Red shorting plug
o
Black dummy plug
o
One primary sensor blanking plug
o
One secondary sensor blanking plug
o
USB with manuals and documentation
o
Battery charger and cable
•
Inspect for damage
o
Check the cable for slices or gouges
o
Check the connector sockets for corrosion, dirt, and salt deposits
o
Check the pressure case for dents and scrapes
o
Check the sensors for cracks or bends
•
Ensure
all the Xchange™
sensors are installed tightly. The blue locking sleeve should be
tight, and sitting less than 1 mm from the instrument end cap.
•
Connect the instrument to a computer with the data cable and perform a scan or monitor if
using SeaCast. Check the battery voltage on the output (normally the last column of data,
unless calculated parameters have been enabled). It should be between 7.2 and 8.4 volts.
Returning an Instrument to the Factory
•
If shipping for repair or recalibration, obtain an RMA number from the service centre.
•
Pack the instrument in its original shipping box to prevent damage during shipping.
An RMA number can be requested using the contact options given in the Support section of this
manual.