5

Installation

3.  Using a stud locator, mark the location of any metal studs in the building wall that intersect the 

headwall mounting location.

stud

drywall

stud reference line

wall

finished floor

4.  Mount the bracket(s) on the wall using hardware compliant to local building codes (supplied 

by others).  Align bracket(s) horizontally according to headwall mounting location.  Refer to 

drawings.  Bracket includes I.D. marks for locating holes vertically, which are to be drilled on-site 

according to location of studs.  Remember to align bottom of bracket to “bottom of bracket” line 

marked in step “2”.

bottom of 

mounting

bracket line

bottom of mounting

bracket lines

bracket

I.D. mark for holes

Summary of Contents for Majestic Series Lite

Page 1: ...Corporation Installation Instructions Horizontal Headwall Majestic Series Lite...

Page 2: ...provided are for the Amico standard products Slight modifications to the installations may be required The product shipped to the job site may not be exactly as shown on the installation drawings Head...

Page 3: ...specific to your installation such as equipment types and quantities room numbers and location shop drawings of each unit type wiring diagrams and rough in locations 2 Locate the headwall unit from th...

Page 4: ...to be installed later This line is to be off set 4 102 mm above devices center line 4 devices center line AFF bottom of mounting bracket line AFF 1 Using shop drawings mark both devices center line AF...

Page 5: ...sing hardware compliant to local building codes supplied by others Align bracket s horizontally according to headwall mounting location Refer to drawings Bracket includes I D marks for locating holes...

Page 6: ...ia panel s for access to rough in locations rough in cutouts 6 Hang the headwall unit on the wall over the mounting bracket and center the unit with refer ence to bed center line Verify rough in locat...

Page 7: ...dwall to the wall through lower section using hardware compliant to local building codes provided by others Lower bracket is to be drilled on site according to stud location Replace cover when done 9...