Operation & Maintenance Manual
Emergency Oxygen Inlet StationSurface Mount High Pressure
Page 1: ...Operation Maintenance Manual Emergency Oxygen Inlet Station Surface Mount High Pressure...
Page 2: ...Contents Introduction 3 Medical Facility Responsibility 3 Installation of Emergency Oxygen Inlet Station 4 Ordering Information 4 Schematic Drawing 4 2 Amico Pipeline...
Page 3: ...on shall be a 1 female npt connection and be provided with a plug to prevent entry of foreign matter into the oxygen system when not in use Amico Emergency Oxygen Inlet Station will perform in conform...
Page 4: ...DED 3 8 16 1 3 4 TAP BOLT 00 M X MAN 49 NIBCO BALL VALVE THREADED 1 00 V X VLV NIB 10T VALVE HANDLE SLEEVE FOR 3 4 TO 1 1 4 00 V C VLV HAND M 3 8 16 HEX NUT 00 H NHF 37NC ELBOW 90 DEG 1 4 1 4 FEM HP 0...
Page 5: ...gen connection 2 NPT 2 Type K Copper Outlet Pipes 9 228 6 6 152 12 305 Main Line Shutoff Valve Pressure Switch Relief Valve 75 PSIG Turn Down and Screen Emergency Low Pressure Oxygen Inlet Box Main Ox...
Page 6: ...ww amico com 85 Fulton Way Richmond Hill Ontario L4B 2N4 Canada Toll Free Tel 1 877 264 2697 Toll Free Fax 1 866 440 4986 Tel 905 764 0800 Fax 905 764 0862 APE OPER MAINT EOISHPSURFMNT V1 0PRINTEDAUG1...