www.amico.com 21
Section 4: Adjustments
2. A sufficient level of tightness may not be achieved by the time the Tilt-setting Lever touches the back of the Patient
Monitor Shelf
(Figure 1)
. If this occurs, PULL the lever AWAY from the MON Head to disengage the lever
(Figure 2)
.
TURN the lever COUNTER-CLOCKWISE
(Figure 3 [1])
to create space for the lever to further tighten. PUSH the lever
TOWARDS the MON Head
(Figure 3 [2])
to re-engage the lever. Repeat Step 1 to further tighten the Tilt Hinge.
3. If a sufficient level of tightness cannot be achieved, use a 4 mm Allen key to loosen the screw that keeps the Tilt-
setting Lever in place
(Figure 4)
. Remove the screw and spring
(Figure 5)
. Remove the lever to reveal the 12 mm
hex-nut
(Figure 6)
.
NOTE:
In some instances (especially with heavy monitors), a sufficient level of tightness cannot be achieved. If
monitor tilt-angle will not set, proceed to Step 3.
4. Turn the 12 mm hex-nut CLOCKWISE using the 12 mm socket wrench
(Figure 7)
to tighten.
5. Test to confirm the tilt-angle is secure.
Figure 4
Figure 6
Figure 7
Figure 5
Spring
Figure 1
Lever blocked by Patient Monitor Shelf
Figure 2
Pull lever AWAY from the MON Head
Figure 3
Rotate Lever COUNTER-CLOCKWISE [1] and PUSH [2]
1
2