background image

AMI Analyzer Manual 

Installation Procedure  

  11 

Calibrating the analyzer 

The oxygen sensor gradually gets used up over its life, and as it does so its sensitivity 

slowly drops until it is at the end of its life, when the sensitivity drops much faster.  

Therefore you have to calibrate the sensor every so often to make sure that your readings 

are correct. 

There are two ways of doing this.  You can use a span gas, a mixed gas containing a 

certain level of oxygen in a background gas that ideally matches your sample, or you can 

use air.  The advantage of using air is that its value really is 20.9% and it doesn’t change; 

the disadvantage is that it means that the sensor is exposed to air and so it will take a 

while for the reading to come down again afterwards.  Typically, at room temperature, 

and AMI analyzer will take about fifteen minutes or less to come down to below 10ppm 

after a one minute exposure to air.   

The advantage of using a span gas is that you can calibrate it to a value close to the 

range of interest – often people use 80ppm oxygen in nitrogen as the span gas.  You may 

have political reasons for having to do this.  The sensor recovers from this level of oxygen 

immediately so there is no down time before the analyzer is working properly again.  The 

disadvantage is that span gases sometimes are incorrectly made, and they can be 

contaminated by improper handling.  Also you will have to provide an external 

sample/span selection valve. 

Calibrating with a span gas 

Fist put a suitable regulator on the span gas tank, and “bleed” it as described below.  

This step is essential, since otherwise the air in the regulator will contaminate the gas in 

the tank.  Connect the regulator to the analyzer with a stainless steel line, and leak 

check the connection with “Snoop™” or similar leak detection fluid.  Purge the line for 

several minutes with a small flow of gas prior to doing this, and leave the gas flowing 

while you make the connection to the analyzer.  This prevents a slug of air from giving 

you excessively high readings when you start spanning the analyzer. 

Make sure the analyzer is seeing a low oxygen level gas – you want the analyzer to go UP 

to the span gas, not down to it, particularly not from air.  Otherwise it will take a very 

long time to get a good calibration. 

When all is assembled, flow the span gas into the analyzer, and you should see the 

reading move to the span gas value.  Assuming it stabilizes somewhere reasonably close, 

press the UP or DOWN button until the reading on the LCD shows what the span gas 

bottle says the value should be.   

Calibrating with air 

You can either connect a compressed air line – from the plant air, not a bottle of 

compressed air – to the sample inlet via a selection valve or you can shut off the flow and 

open the cell cap on the front of the analyzer.  If you use compressed air, go through the 

same procedure as above, only adjust the oxygen reading to 20.9%. 

If you do not have compressed air, unscrew the cell cap on the front of the analyzer.  

Blow some air under the sensor by waving a book or some such at it. Adjust the reading 

as before to 20.9% and then screw the cell cap back on the analyzer.  Bring up the flow of 

the sample gas. 

In either case, make sure you don’t take more than a minute to do this.  The reading may 

not stabilize exactly at 20.9%, but don’t worry about that – any slight error will be 

inconsequential at the operating levels. 

Let the analyzer come down to a low reading on the sample gas. 

Summary of Contents for 2001LC

Page 1: ...AMI Oxygen Analyzer Manual Model 2001LC AMI Costa Mesa CA ...

Page 2: ...ion and Operation 4 Receiving the analyzer 4 Installation 4 Location 4 Precaution 4 Installation Procedure 5 Connect gas and power lines 7 Interconnections 8 Alarm connections 8 Output connections 8 Serial connections 8 Sample Handling 9 Operation 10 Analyzer operation 10 Sample tubing and components 10 Leaks 10 Exhaust 10 Calibrating the analyzer 11 Output range concept 12 Front Panel Controls 12...

Page 3: ...Bypass 14 Verify Span Factor 14 Read the Temperature 14 Alarm Functionality 14 Maintenance and troubleshooting 15 Maintenance 15 Periodic Calibration 15 Sensor Replacement 15 Sensor replacement cautions 16 Sensor replacement procedure 17 Bleeding a regulator 18 Troubleshooting 19 All oxygen applications 19 Specifications and Disclaimer 21 Specifications 21 Disclaimer 22 ...

Page 4: ...the air contains 20 9 209 000ppm of oxygen and it can get into a pressurized pipeline through the smallest leaks Oxygen molecules will enter through a leak no matter the pressure or the nature of the gas in the line This analyzer is an exceptionally sensitive leak detector including those provided by improper installation Make sure you read this manual carefully prior to installation Caution Read ...

Page 5: ...us area This manual covers software version 1 0 Features Compact size Unique patented cell block Auto ranging display with user selectable output range Front panel sensor access Optional air or span gas calibration no zero gases required Virtually unaffected by hydrocarbons or other oxidizable gases High accuracy and fast response Large liquid crystal display Backed by a two year warranty excludin...

Page 6: ...h much faster come down times typically under twenty minutes to 10ppm from a 1 minute air exposure Sensor Warranty The sensor is warranted to operate for a period determined by its class If the sensor ceases to operate correctly before this time has elapsed contact AMI for a return authorization for evaluation If there is any evidence of defective material or workmanship the sensor will be replace...

Page 7: ...panel in a general purpose area It should be mounted at a suitable viewing level Refer to the drawing figure 1 showing the analyzer dimensions It is not suitable for use in a hazardous area or with flammable gases Although the unit is RFI protected do not to mount it close to sources of electrical interference such as large transformers motor start contactors relays etc Also avoid subjecting it to...

Page 8: ... 3 Deal with any potential condensation or liquid contamination issues 4 Connect the sample line to the sample inlet port with ss tubing 5 Pressurize the sample line to line pressure between 1psig and 20psig 6 Leak check every fitting and weld from the analyzer inlet to the sample tap 7 Connect vent line to outside or a suitable purge system 8 Connect power relay contacts and analog output If usin...

Page 9: ... spray e Disconnect the sample tubing from the inlet port either by physically removing it or by using an external three way valve f Flow calibration gas WHILE you are connecting the span gas tubing to the inlet fitting or the selection gas valve port Allow the gas to purge through the fitting for about 20 seconds before you tighten it g Press the ALARM BYPASS button and adjust the time displayed ...

Page 10: ...to a suitable vent Sample gas The sample gas inlet pressure should be between 1 40psig Span gas Span gas if desired must be provided by a user supplied valve Exhaust The exhaust line may be left open or vented to a suitable vent If used with a scavenging system use a large diameter pipe for example pipe as the input to the scavenging system and allow the vent to terminate a little way inside this ...

Page 11: ...By failsafe is meant that the relays are powered when NOT in alarm so that if power fails they indicate an alarm condition The contacts can handle AC or DC voltages and can carry up to 5A of current for a resistive load Inductive loads such as solenoid valves should be snubbed we suggest that you connect diodes or Zener diodes or Transzorbs directly across them to absorb the inductive spike Do not...

Page 12: ...ure within this range Higher pressure applications or varying pressure applications will need a regulator to control pressure The analyzer is not sensitive to flow changes between about 0 2SCFH and 5 SCFH but it is sensitive to back pressure changes changes in exhaust pressure The flowmeter is mounted on the exhaust so that the oxygen reading is not affected by potential leaks around the flowmeter...

Page 13: ... through its O rings The span gas pressure should also be held below 100psig If you permanently connect a span gas tank to it via a user supplied three way valve make very sure that the connection is leak tight as otherwise not only will the calibration be in error but also you will lose all of your span gas as it leaks out Leaks If you suspect a leak due to high oxygen readings change the flow ra...

Page 14: ...will contaminate the gas in the tank Connect the regulator to the analyzer with a stainless steel line and leak check the connection with Snoop or similar leak detection fluid Purge the line for several minutes with a small flow of gas prior to doing this and leave the gas flowing while you make the connection to the analyzer This prevents a slug of air from giving you excessively high readings wh...

Page 15: ...ecause that air calibration is in fact very valid You will have to trouble shoot whatever has happened with your gas Output range concept The analyzer displays the oxygen level in appropriate units on the LCD automatically adjusting its sensitivity as required Meanwhile the analog output and the alarms are set on a single user selectable Output range For example you can set the analog output to co...

Page 16: ...n If the output range does not change the security level must be set to full or span only security In this case change the security level with the laptop and the AMI User Interface program Alarm Set Points The alarm set points can be viewed and changed from the front panel View Alarm Set Points Press either of the ALARM SET POINT buttons and let go The alarm set point will be displayed for about 3...

Page 17: ... up slowly over the life of the sensor When the value has gotten up to 1000 the sensor has reached the end of its life and should be replaced Also if the value suddenly jumps it indicates that the sensor is getting close to the end of its life Read the Temperature Press the DOWN arrow button The display will show the temperature of the cell block in degrees Fahrenheit The value is limited to 25F a...

Page 18: ...bration is usually satisfactory Use in a particularly aggressive environment may degrade the sensor faster in this case calibrate more often Sensor Replacement This should be done based on the Span Factor feature rather than as a response to a dead sensor See the chart below for recommended sensor replacement Sensor Part number Description Expected life T2 4SEN09 1 Trace oxygen CO2 background 9 mo...

Page 19: ... it does immediately flush the affected area with water for a period of at least 15 minutes Refer to the Material Safety Data Sheet provided Dispose of leaking or used sensors in accordance with local regulations Sensors usually contain lead which is toxic and should generally not be thrown into ordinary trash Refer to the MSDS to learn about potential hazards and corrective actions in case of any...

Page 20: ...lly open the bag using a pair of scissors or a knife Make sure you don t cut yourself or stab the sensor In the rare event that the sensor has leaked there will be liquid in the bag If so do not proceed you need a new sensor Be careful that you don t poke anything such as a fingernail through the membrane 8 Don t pull out the shorting tab yet 9 Holding the sensor by its tab membrane side down slid...

Page 21: ... to Bleed the regulator first 1 Install the regulator on the span gas bottle but do not open the bottle valve yet 2 Close the regulator exit valve 3 Close the regulator to what would be no pressure on the outlet 4 Briefly open the bottle valve and close it again 5 Loosen the nut connecting the regulator to the bottle and allow the pressure to bleed off and then tighten it again 6 Open the regulato...

Page 22: ... soap solution or equivalent 4 Verify that the gas flow rate is correct 0 1 to 2 SCFH 5 Oxygen diffusion can be a serious problem Verify that no plastic tubing or other plastic components are used in a trace gas system including diaphragms of pressure regulators packing of valves etc For percent applications similar problems may be experienced with silicone tubing Use Teflon or Tygon or similar hi...

Page 23: ... sensor replacement instructions under Maintenance No voltage or current output to recording device 1 Verify that the output wires are properly stripped and connected 2 Verify the connections on the output terminal block 3 Verify that the output connections are not shorted all the way back to the recording device Disconnect the wires from the analyzer and use an ohmmeter to check for shorts or ope...

Page 24: ...1 psig to 100psig Humidity 95 non condensing Operational conditions Pollution degree 2 Installation category I I Drift 1 of full scale in 4 weeks at constant temperature dependent on sensor Expected cell life 9 months to 2 years Response times 90 of full scale in less than 0 10 ppm 25 sec 0 100 ppm 10 sec 0 1000 ppm 10 sec Output 4 20mA isolated Alarm contacts 230 117VAC 5A or 28VDC 5A resistive P...

Page 25: ...s or associated components AMI s obligation is expressly limited to the analyzer itself The AMI analyzer is not designed as a primary safety device that is to say it is not to be used as the primary means of assuring personnel safety In particular it is not designed to act as a medical instrument monitoring breathing air for correct oxygen concentration and should not be used as such when it is th...

Reviews: