WARNING
The absolute maximum voltage for digital counter channel’s is limited to 25 V. Beyond 25 V, the
EX1200-1538 will be permanently damaged. Additionally, the inputs should never be driven
below the ground potential or damage may occur.
The DIO channels allow for the maximum voltage of 60 V per channel. When configured as a
digital output channel, the current flowing in either direction should be limited to 50 mA.
Analog outputs have internal open and short circuit protected. Any external voltage applied should