System Overview
|
2-13
Electronics
• One assembly - mounted to a plate
• The electronics/software completely control the sensor.
- No external controllers are required.
• The configuration/monitoring interface to the electronics is RS485
MODBUS RTU.
• Sensor wiring and Customer wiring (e.g., power, I/O) are directly con-
nected to the main electronics.
- Connections are pluggable.
- Both plugs and headers are laser printed for easy matching.
I/O
The WDG-V has the following I/O available:
• Qty (3) configurable 4-20 mA outputs
• Qty (2) system alarm relays (Service and Data Valid)
• Qty (3) configurable process alarm relays
• Qty (1) RS485, 2 wire, MODBUS RTU interface (57.6 Kbaud)
• Qty (1) Contact input (for remote auto calibration trigger)
• Valve controls for Remote Calibration Unit (RCU)
Analog Outputs
There are 3 analog outputs in the WDG-V. Each analog output can be con-
figured for Mode, Function, and Span, Zero.
Mode options include: 0-20ma, 4-20ma and NAMUR.
Functions include: Oxygen, cell millivolts, combustibles*, methane*
The analog outputs can be powered internally or externally. External
power must be greater than 26V.
The NAMUR outputs are implemented as follows:
4-20 mA Signal
Analyzer Condition
0 mA
Analyzer unpowered, or completely failed
3.5 mA
Critical Alarm - analyzer reading unusable (factory default)
3.8 mA
Reading Under Range (Example: user sets range to 2-10%. Current reading is 1.9%)
4 to 20 mA
Normal Operation
20.5 mA
Reading Over Range (Example: range is 0-10%. Current reading is 12%)
* For combustible and/or methane units only.
Summary of Contents for Thermox WDG-V
Page 8: ...viii Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 16: ...xvi Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 22: ...1 6 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 34: ...2 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 57: ...Installation Guide 3 17 Figure 3 18 RCU flow diagram 3 Span Combustibles and Methane...
Page 58: ...3 18 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 64: ...3 24 Thermox WDG V VC VCM with Blow Back Figure 3 20 Wiring...
Page 73: ...AMEVision Display User Interface or Communications 4 5...
Page 76: ...4 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 88: ...5 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 91: ...Service and Parts 6 3 Figure 8 1a Wiring diagram for sensor...
Page 96: ...6 8 Thermox WDG V VC VCM with Blow Back Figure 8 3 Thermocouple replacement...
Page 100: ...6 12 Thermox WDG V VC VCM with Blow Back Figure 8 5 Flow sensor replacement...
Page 103: ...Service and Parts 6 15 Figure 8 7 Box Heater Replacement...
Page 105: ...Service and Parts 6 17 Figure 8 8 Combustible Block Heater Replacement...
Page 107: ...Service and Parts 6 19 Figure 8 9 Electronics Board Removal...
Page 114: ...7 4 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 118: ...A 4 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 126: ...A 12 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 132: ...A 18 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 137: ...PC Configurator Software A 23 Figure 23 Calibration_Calibration Results screen...
Page 138: ...A 24 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 144: ...A 30 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 156: ...A 42 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 164: ...A 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...