SECTION 3

 

THEORY OF OPERATION 

3.1 General 

The M51 interface board consists of two functional blocks, the isolation amplifier with input 
buffer amplifier and associated switching circuitry, and the isolated power supply, which 
provides power to the input side of the isolation amplifier.   

3.2 Isolation 

Amplifier 

The isolation amplifier U1, has a capacitively isolated input stage, which prevents any D.C. 
currents from flowing between the programming source, and the power supply.  It has a fixed 
overall unity gain, which allows for simple interfacing of the external program signal and the 
internal control points.  U3 amplifies and buffers the input signal, with R8 and R11 providing 
range and offset adjustment points, while R7, R6 and R5 in conjunction with SW1-7 and 8 set 
the appropriate program scale. 

Quad analog switch, U2, connects the output of the amplifier to either the voltage program line 
(J2-9) or the current program line (J2-8), depending on the setting of SW1-2 and SW1-3 (see 
Table 2-1 for settings).  The control line that is not being programmed is connected to the R4, 
R3 voltage divider, which provides a fixed full scale (5V) program signal.  SW1-1 sets the power 
supply in the remote control mode by pulling pin J2-4 low.  In this way the single isolation 
amplifier can be used to program either the output voltage or current, while the other non-
programmed line is held at the full scale level, independent of where the front panel control 
knobs are set. 

3.3  Isolated Power Supply 

The isolated power supply is required to p12V and -5V to the input side of the isolation 
amplifier, while maintaining electrical isolation between this section of the M51 board and the 
DCS unit output. 

The power supply is a low power flyback type circuit, controlled by U5, a 100kHz current mode 
pulse width modulator (PWM).  This PWM controls the on time of the switching transistor Q1, a 
N-channel MOSFET, which controls the power delivered to the secondary T1, the isolation 
transformer.  The transformer winding voltage is sensed at the control circuit by rectifying and 
filtering the output of the control winding at T1 pin 9 and 2.  This voltage is divided by R13 and 

M360002 Rev D 

3-1 

Summary of Contents for Sorensen DCS M51 Series

Page 1: ...M360002 Rev D www programmablepower com DCS Series M51 Isolated Analog Option Programming Manual...

Page 2: ......

Page 3: ...ice of Copyright DCS Series M51 Isolated Analog Option Programming Manual 2002 2009 AMETEK Programmable Power Inc All rights reserved Exclusion for Documentation UNLESS SPECIFICALLY AGREED TO IN WRITI...

Page 4: ...This page intentionally left blank ii...

Page 5: ...and equipment safety Always ensure that facility AC input power is de energized prior to connecting or disconnecting any cable In normal operation the operator does not have access to hazardous volta...

Page 6: ...This page intentionally left blank iv...

Page 7: ...OPERATING INSTRUCTIONS 2 1 2 1 Programming Connection 2 1 2 2 Programming Mode 2 1 2 3 Operation 2 2 2 4 Calibration 2 2 SECTION 3 THEORY OF OPERATION 3 1 3 1 General 3 1 3 2 Isolation Amplifier 3 1 3...

Page 8: ...M51 Isolated Analog Option DCS Series This page intentionally left blank vi M360002 Rev D...

Page 9: ...ndard 15 pin female D SUB connector 1 2 Specifications Maximum Voltage Differential Between program input power supply return 600VDC Remote Programming Control Coefficient 1 Output Voltage 0 10V pgm S...

Page 10: ...M51 Isolated Analog Option DCS Series This page intentionally left blank 1 2 M360002 Rev D...

Page 11: ...ge or the output current of the power supply or it may be set for standard local front panel control The input can also be configured to accept either a 0 10V 0 5V or 0 400mV control signal The operat...

Page 12: ...he programming voltage CAUTION Never allow the power supply output voltage or current to exceed the rated maximum 2 4 Calibration If the program range or offset should require calibration adjustments...

Page 13: ...e 2 1 for settings The control line that is not being programmed is connected to the R4 R3 voltage divider which provides a fixed full scale 5V program signal SW1 1 sets the power supply in the remote...

Page 14: ...7 and 8 regulated The main secondaries are rectified and filtered to provide approximately 15V to the post regulators U3 and U4 U3 provides a fixed 12V to the input stage of U1 while U4 provides a fi...

Reviews: